Alejandra Gonzalez-Calle, Runze Li, Isaac Asante, Juan Carlos Martinez-Camarillo, Stan Louie, Qifa Zhou, Mark S. Humayun, "Development of Moderate Intensity Focused Ultrasound (MIFU) for Ocular Drug Delivery", BME Frontiers, vol. 2022, Article ID 9840678, 8 pages, 2022. https://doi.org/10.34133/2022/9840678

Development of Moderate Intensity Focused Ultrasound (MIFU) for Ocular Drug Delivery

Received30 Nov 2021
Accepted22 Mar 2022
Published17 Jun 2022

Abstract

The purpose of this study is to develop a method for delivering antiinflammatory agents of high molecular weight (e.g., Avastin) into the posterior segment that does not require injections into the eye (i.e., intravitreal injections; IVT). Diseases affecting the posterior segment of the eye are currently treated with monthly to bimonthly intravitreal injections, which can predispose patients to severe albeit rare complications like endophthalmitis, retinal detachment, traumatic cataract, and/or increased intraocular. In this study, we show that one time moderate intensity focused ultrasound (MIFU) treatment can facilitate the penetration of large molecules across the scleral barrier, showing promising evidence that this is a viable method to deliver high molecular weight medications not invasively. To validate the efficacy of the drug delivery system, IVT injections of vascular endothelial growth factor (VEGF) were used to create an animal model of retinopathy. The creation of this model allowed us to test anti-VEGF medications and evaluate the efficacy of the treatment. In vivo testing showed that animals treated with our MIFU device improved on the retinal tortuosity and clinical dilation compared to the control group while evaluating fluorescein angiogram (FA) Images.

1. Introduction

Diseases affecting the posterior segment of the eye, such as age-related macular degeneration (AMD), diabetic retinopathy, retinal vein occlusion, and posterior uveitis, are difficult to treat because of the challenge of delivering therapeutically relevant concentrations of medication into the vitreous humor (VH) and retinal tissue [1]. Although topical eye drops are safe, this route of delivery is incapable of achieving therapeutic levels in the human VH required to exert their pharmacologic activity [2]. To ensure therapeutic concentrations, most drugs used to treat retinal conditions such as AvastinTM, LucentisTM and EyleaTM are directly injected into the posterior segment as an intravitreal (IVT) injection [3]. To maintain therapeutic concentrations, these agents must be given monthly or bimonthly. However, frequent IVT injections can predispose these patients to rare but severe complications like endophthalmitis, retinal detachment, increased intraocular pressure, cataractogenesis, and increased retinal ganglion cell apoptosis [4]. Thus, the ability to noninvasively deliver effective agents may reduce potential complications while improving treatment adherence.

Transscleral delivery is a novel strategy to efficiently achieve therapeutic level of drugs in the retina and choroid tissue by exploiting the large scleral surface area [5]. Despite the intrinsic permeability characteristics of the sclera, it continues to be a challenge to deliver high molecular weight (MW >70 KDa) compounds into the posterior segment [1]. Efforts to exploit this strategy include transscleral iontophoresis [6], which uses an electric field to deliver drugs into the retina and choroidal tissue. However, iontophoresis may be limited to charged drugs. Unfortunately, this procedure is not well tolerated by patients. Therefore, there is a need to develop a delivery system that can be used across a large spectrum of therapeutic agents and that is independent of its molecular weight.

In this paper, we developed a method for delivering antiinflammatory agents of high molecular weight (e.g., Avastin) into the posterior segment that is less invasive than IVT injections. Our novel delivery method employs moderate intensity focused ultrasound [MIFU]) to mediate scleral thinning and thus enhance drug transport across this barrier. To evaluate the efficacy of our delivery method, we developed a rabbit disease model using vascular endothelial growth factor (VEGF-A 165) to replicate the neovascular retinopathy.

2. Results

2.1. Transducer Performance Evaluation
2.1.1. Basic Performance Evaluation

The center frequency of the transducer was measured as 3.27 MHz with the -6 dB bandwidth of 9%. The results indicated that it reached a good agreement with PiezoCAD simulation, where the center frequency was 3.25 MHz and the -6 dB bandwidth was 16% (Figures 1(a) and 1(b)).

Hydrophone (HGL-0400, ONDA, CA, USA) measurement was conducted to characterize the acoustic field of the transducer (Figure 2). The acoustic pressure distribution was acquired as well as the lateral and axial beamwidth. The measured lateral and axial beamwidth were 0.663 mm and 1.872 mm, respectively. Finite-element analysis was conducted by COMSOL; the lateral and axial beamwidth was 0.778 mm and 1.554 mm, respectively.

2.1.2. Thermal Effect Evaluation

The thermal effect caused by the interaction of the ultrasound and the sclera was measured with a needle-type thermocouple. The results showed that the temperature was increased 1.8 °C in total (Figure 1(c)).

2.1.3. Cavitation Effect Evaluation

The most commonly used passive cavitation detection (PCD) method was selected to evaluate the cavitation effect. The PCD system mainly contained 2 parts: detection transducer and emit transducer. The detection transducer had a center frequency of 15 MHz with a -6 dB bandwidth of 60%, and the sensitive frequency range of 11 MHz to 19 MHz was selected in this study. It was aligned confocally with the emit transducer guided by the hydrophone, and the sclera was placed in this focal spot. The detection transducer received the reflected ultrasound signal from the sclera. If the cavitation effect occurred, then a broadband shock wave would be generated due to the bubble collapsing. The cavitation effect was assumed to have occurred when the broadband noise caused by the shock wave in the frequency range mentioned above had more than √5 of amplitude comparing with the background signal [7, 8]. In Figure 3, despite the subharmonic signal at 16.35 MHz, the measured results met this criterion (Figure 1(d)).

2.2. Experimental Study
2.2.1. MIFU Treatment

MIFU treatment was applied at 3 different durations (1 min, 2 min, and 3 min) to evaluate and observe the changes caused in the sclera. Scleral thinning was observed for all three durations (Figure 4(b)). Since the objective of our study is to enhance large molecular therapeutic such as Avastin to transit through the barrier, it was determined qualitatively that 3 min MIFU treatment could maximally thin the sclera without disrupting structural integrity.

2.2.2. Treatment Groups

Retinal vascularization in all animals was scored using Figure 5. A severity scoring system from 0 to 4 was used to quantify signs of tortuosity and neovascularization, where 4 was defined as the most severe and 0 as normal. Vascularization was observed in all the rabbits at day 3 after VEGF induced CNV (score: ). The injected eye showed signs of vasodilation and tortuosity and neovascularization (Figures 3 and 6). Rabbits in group 2 received an IVT injection of 1.25 mg Avastin 3 days after VEGF injection. 4 days after the avastin injection, animals showed a clinical and angiographic improvement (Score: 0 ± 0). On the contrary, animals from group 3 which received a subconjunctival injection of 1.25 mg Avastin 3 days after VEGF injection (Score: 3.33 ± 0.58) did not show retinal vascular changes 4 days post-avastin (3.33 ± 0.58). However, in NZB, receiving both MIFU treatment plus a subconjunctival injection of 1.25 mg Avastin showed an improvement on the retinal tortuosity and clinical dilation compared to their findings at day 3 (score at day 3: ); FA images showed mild dilation with minimum leakage (score 4 days post Avastin: ). Thus, group 4 shows a similar clinical effect compared to group 2. All animals from three different groups were scored using Figure 5 using a 0 to 4 severity scoring system, being 4 the most severe and 0 normal, and the results are summarized in Table 1.


ScoringGroup 2:
intravitreal injection
Group 3:
subconjunctival injection alone
Group 4:
MIFU + subconjunctival injection

Day 0
Day 3 post-VEGF injection
Day 7–day 4 post-Avastin

3. Conclusion and Discussion

In this study, a highly focused transducer was fabricated and utilized to mediate thinning of the sclera to enhance drug transport across this barrier. Being able to create a highly focused transducer allowed us to have a precise and focalized window of treatment. For safety reasons, it was important to validate that the thinning of the sclera was caused by the cavitation effect instead of the thermal effect. Temperature changes were evaluated during a three-minute trial of continued pulses, and an increase in temperature of 1.8 °C was observed during that time indicating a minimal thermal effect. Thus, it can be concluded that the thinning of the sclera was predominantly caused by the cavitation effect.

The cavitation bio effects can be quantified by means of a unitless measurement known as the mechanical index. In order to assess safety, the FDA has stipulated that the mechanical index of ultrasound devices used in the ophthalmology field for diagnostic purposes cannot exceed a value of 0.23. Although our current device has a mechanical index of 10, our study showed no adverse effects on other tissue structures besides the thinning of the sclera observed in the OCT images. Since our ultrasound device is meant to be used for treatment purposes, the FDA proposed safety limit does not directly translate to our application. Future studies should focus on finding mechanical index limits for the safe utilization of ultrasound with treatment purposes.

The high intensity focused ultrasound (HIFU) has been used by other groups in the ophthalmology field with treatment purposes in both clinical and preclinical applications [9, 10], and safety issues have rarely been reported [11]. For example, Razavi et al. used an ultrasound device with mechanical index ranging from 6.5 to 12.2 to modify the permeability of the sclera. This work showed that inertial cavitation enhances the permeability of the sclera with no significant alterations of the tissue observed in the electron microscopy images [12]. Aptel et al. utilized ultrasound coagulation of the ciliary body with HIFU to reduce intraocular pressure in both an animal study and a pilot human study. Their results showed that HIFU was well tolerated and no significant damage was observed in the histology images in the animal study [13]. Similarly, no major intraoperative or postoperative complications were reported in the human study [14]. The aforementioned results provide evidence that focused ultrasound can be used safely for treatment purposes in ophthalmology applications.

To test the efficacy of this device in vivo, we were able to successfully create a disease model using pigmented rabbits to mimic the behavior of retinal retinopathies by using VEGF-165. Other groups have also developed models with persistent retinal neovascularization by using either Dutch belted rabbits [15] or New Zealand rabbits [16] . This paper on the other hand created a retinal disease model that could be used for both breeds allowing us to compare the behavior of the VEGF-165A based on the different breed characteristics. This information will be useful for future research that aims at testing new protocols to treat retinopathies.

By using a disease model instead of normal pigmented rabbits, we were able to test the efficacy of the novel drug delivery system by objectively quantifying the Anti-VEGF medication (Avastin) effects on the retina. After comparing the control (subconjunctival injection only) versus the MIFU treated group (MIFU + subconjunctival injection) based on the FA images, we observed a definitive improvement in the tortuosity and neovascularization of the retina in the MIFU group, while no improvement was observed in the control group. This allowed us to conclude that sufficient concentration of medication penetrated the sclera when the MIFU device was used.

We have shown that the scleral thinning obtained by the MIFU treatment can facilitate the penetration of large molecules across the scleral barrier. Our pilot study has demonstrated that there is a negative correlation between the permeability of the sclera and overall thickness. Despite the fact that the rabbits’ sclera is thinner compared to humans’ sclera, we believe that the observed correlation may still apply to humans. This work has shown promising evidence that this is a viable method to deliver high molecular weight medications not invasively. It should be noted that our in vivo results are based on a small sample size and only assessed the thinning of the sclera immediately after treatment. Our future work will focus on further testing safety and efficacy of our device in a larger sample size and studying the long term permeability of the treated area for transscleral transport, since currently, patients receive multiple Avastin treatments for long periods of time.

4. Materials and Methods

4.1. Transducer
4.1.1. Experimental Setup

For the in vivo studies, a PZT-4 based transducer with an aperture size of 20 mm and the focal length of 13 mm was fabricated. Parylene C (Specialty Coating Systems, Indianapolis, IN, USA) was used to compensate for the acoustic mismatch between the piezoelectrical material and the water, and it also served as a water insulation layer [17]. Air was used as the backing material to enable max acoustic power output [18].

A water-cooling system circulates water through the transducer housing, removing the heat when the transducer is working, allowing us to use it for longer periods of time during in vivo study. The focused ultrasound passes through the small hole at the bottom of the water cone allowing us to focus accurately on the sclera (focal spot) (Figure 4(a)).

The transducer was held by a five-axis translational stage to enable a perpendicular ultrasonic beam to the sclera. The transducer was driven by a function generator (AFG 3252 C, Tektronix, Beaverton, OR, USA) and a RF amplifier (100A250A, Amplifier Research, PA, USA) with 55 dB of gain. The ultrasonic transducer was placed on the focal spot. In order to select the optimal stimulation time, various pulse durations (1, 2, and 3 min) were evaluated in vivo using OCT images (Figure 4(b)). The size and depth of the scleral thinned area after treatment were used to select the ideal stimulation duration. The mechanical index was set around 10 with duty cycle of 12%. Table 2 summarizes the waveform generator parameters used during MIFU stimulation.


Waveform generator parameters
WaveformSinewave

Peak to peak voltage0.4Vpp
Duty cycle400 cycle: burst
Pulse period1 ms burst period
Output frequency3.3 MHz

4.1.2. Performance Evaluation Tests

To evaluate the performance of the transducer, several characterizations have been made as followed: pulse-echo test and PiezoCAD simulation, hydrophone test and COMOSOL (COMSOL 5.3a, COMSOL Inc., MA, USA) simulation, thermal effect, and cavitation effect mechanism investigations. Pulse-echo measurement was described in our previous work with the same procedures [19], and hydrophone measurement was also provided in our previous work [20]. The investigation of the thermal effect was achieved by a thermal couple. The sclera was positioned on the focal spot of the transducer, and a lead was inserted under the sclera shell. Every 0.1 °C of temperature raised, the corresponding time point was recorded, and the total time span was 3 min. The cavitation effect was conducted by a confocally transmit transducer and a receive transducer. The received transducer recorded the reflected echo from the sclera; the echo signal was saved for offline processing.

4.2. Animal Model

Rabbits share anatomical and physiological similarities with the human eye. Ocular size, volume, internal features, and their diffusional path to reach the anterior segment using topical administered compounds are some of them, making it a great model for our purpose [21].

All animals were maintained on a daily 12 hr light/dark cycle. All procedures were in conformance with the Guide for Care and Use of Laboratory Animals (National Institutes of Health). The University of Southern California Institutional Animal Care and Use Committee (IACUC) reviewed and approved all procedures before commencement of the study.

We performed survival experiments during our study. Adult New Zealand Black (NZB) rabbits (Western Oregon), ~2-3 kg, were used (). All experiments were performed in the right eye of each animal. The animals were anesthetized with ketamine (100 mg/kg) and xylazine (20 mg/kg) combination and euthanized at the end of the study.

4.2.1. Disease Model

Retinopathies are characterized by the development of choroidal neovascularization (CNV) through endogenous activation of VEGF elaboration [22, 23]. To validate the efficacy of the drug delivery system, a disease model replicating the behavior of retinal retinopathies was created. The creation of this model allowed us to test anti-VEGF medications and evaluate the efficacy of the treatment.

To find the correct concentration of VEGF-A 165 that is needed to create our disease model, rabbits from two different breeds (New Zealand black and pigmented Dutch belted) were stratified into two different groups to receive either 1 μg or 2 μg of freshly reconstituted VEGF-A 165 intravitreally. Ophthalmologic examinations were conducted on days 7 and 14 to determine the severity as well as the resolution of the retinal edema and angiogenesis at designated time points. At the end of the study (EOS) or day 14, the fundus and fluorescein angiography (FA) images were scored by a “Masked” evaluator for signs of tortuosity and neovascularization using a 0 to 4 severity scoring system (Figure 5).

4.3. Experimental Study

The eye was dilated with two drops of 1% tropicamide (Tropicacyl, Akorn Inc., Buffalo Grove, IL) and 2.5% phenylephrine (AK-Dilate, Akorn Inc., Buffalo Grove, IL). The animal’s heart rate, blood pressure, temperature (rectal), and respiratory rate were monitored and recorded every 5 minutes during surgical procedure and recovery time for survival experiments. The animal’s body temperature was maintained at 37 °𝐶 with an electric heating pad.

The VEGF-A 165 (R&D systems; catalogue number 293-VE-050/CF) was freshly reconstituted at a concentration of 3 μg/50 μL. On day 1, baseline ocular examination, fundoscopic images, and fluorescein (FA) image were taken. The rabbits received 50 μl as an intravitreal injection of VEGF-A into the right eye.

On day 3, ophthalmologic examinations using funduscope and fluorescein angiography (FA) were used to assess the severity of the angiogenesis and retinal edema (Figure 5). The transducer was held by a three-axis translational stage (Model 4044 M Parker Daedal, Cleveland, OH) mounted on a magnetic based articulating arm for precise placement. The transducer was placed on the surface of the sclera ~3 mm from the limbus without applying any pressure to it. On day 7, at the conclusion of the study, all animals were euthanized with an intravenous injection of pentobarbital (30 mg/kg; Butler, Dublin, OH).

4.3.1. Treatment Groups

Four treatments groups were studied (Table 3). The fundus images and fluorescein images were acquired for all groups at day 0, day 3, and day 7.


IDTreatmentNVEGFAvastinMIFUEOS

ControlVEGF (3 μg)3D0NANAD7
Control 1VEGF (3 μg) + Avastin (1.25 mg) IVT3D0D3NAD7
Control 2VEGF (3 μg) + Avastin (1.25 mg) SC3D0D3NAD7
TreatmentVEGF(3 μg) + MIFU(3 min) + Avastin(1.25 mg) SC3D0D3D3D7

(1) Group 1: VEGF Control. 3 rabbits received an intravitreal injection of 3 μg/50 μL of VEGF into the right eye and followed up at day 3 and day 7 to study the VEGF behavior without treatment.

(2) Group 2: VEGF + Intravitreal (IVT) Avastin Only. At day 0, 3 rabbits received an intravitreal injection of 3 μg/50 μL of VEGF into the right eye. At day 3, all animals received an intravitreal injection of Avastin (1.25 mg/50 μL) in the treated eye. At day 7, ophthalmologic examinations using funduscope and fluorescein angiography (FA) were used to assess control treatment group.

(3) Group 3: VEGF + Subconjunctival Avastin Only. At day 0, 3 rabbits received an intravitreal injection of 3 μg/50 μL of VEGF into the right eye. At day 3, all animals received a subconjunctival injection of Avastin (1.25 mg/50 μL) in the treated eye. At day 7, ophthalmologic examinations using funduscope and fluorescein angiography (FA) were used to assess control treatment.

(4) Group 4: VEGF + MIFU + Subconjunctival Avastin. At day 0, 3 rabbits received an intravitreal injection of 3 μg/50 μL of VEGF into right eye. At day 3, all animals received MIFU treatment for 3 minutes (settings summarized in Table 1) and a subconjunctival injection of Avastin (1.25 mg/50 μL) in the right eye on top of treated area. At day 7, ophthalmologic examinations using funduscope and fluorescein angiography (FA) were used to assess treatment.

Data Availability

Data is available upon request from the authors.

Conflicts of Interest

The authors declare that there is no conflict of interest.

Authors’ Contributions

AGC conducted the animal experiments, wrote the manuscript, and prepared the figures. RL built the transducer, conducted transducer testing, wrote the manuscript, and prepared the figures. IA developed the rabbit disease model. JCMC performed the surgical procedures. SL, QZ, and MH supervised the project and edited the manuscript. Alejandra Gonzalez-Calle and Runze Li contributed equally to this work.

Acknowledgments

This study was supported by the USC Steven Technology Advancement grant and the Jean Perkins Foundation Funding.

References

  1. P. M. Hughes, O. Olejnik, J. Chang-lin, and C. G. Wilson, “Topical and systemic drug delivery to the posterior segments,” Advanced Drug Delivery Reviews, vol. 57, no. 14, pp. 2010–2032, 2005. View at: Publisher Site | Google Scholar
  2. H. Nomoto, F. Shiraga, N. Kuno et al., “Pharmacokinetics of bevacizumab after topical , subconjunctival, and intravitreal administration in rabbits,” IOVS, vol. 50, no. 10, pp. 4807–4813, 2009. View at: Google Scholar
  3. M. Kataja, P. Hujanen, H. Huhtala, K. Kaarniranta, A. Tuulonen, and H. Uusitalo-jarvinen, “Outcome of anti-vascular endothelial growth factor therapy for neovascular age-related macular degeneration in real-life setting,” Journal of Ophthalmology, vol. 102, no. 7, p. 965, 2018. View at: Publisher Site | Google Scholar
  4. R. D. Jager, L. P. Aiello, S. C. Patel, and E. T. Cunningham, “Risks of intravitreous injection: a comprenhensive review,” Retina, vol. 24, no. 5, pp. 676–698, 2004. View at: Publisher Site | Google Scholar
  5. G. A. Rodrigues, D. Lutz, J. Shen et al., “Topical drug delivery to the posterior segment of the eye : addressing the challenge of preclinical to clinical translation,” Pharmaceutical Research, vol. 35, no. 12, pp. 1–5, 2018. View at: Publisher Site | Google Scholar
  6. S. A. Molokhia, E. Jeong, W. I. Higuchi, and S. K. Li, “Examination of barriers and barrier alteration in transscleral iontophoresis,” Journal of Pharmaceutical Sciences, vol. 97, no. 2, pp. 831–844, 2008. View at: Publisher Site | Google Scholar
  7. T. Li, H. Chen, T. Khokhlova et al., “Passive cavitation detection during pulsed Hifu exposures of ex vivo tissues and in vivo mouse pancreatic tumors,” Ultrasound in Medicine and Biology, vol. 40, no. 7, pp. 1523–1534, 2014. View at: Publisher Site | Google Scholar
  8. R. Alberts, Vision: Human and Electronics, Springer, 1st edition, 1973.
  9. S. E. P. Burgess, R. H. Silverman, D. J. Coleman et al., “Treatment of glaucoma with high-intensity focused ultrasound,” Ophthalmology, vol. 93, no. 6, pp. 831–838, 1986. View at: Publisher Site | Google Scholar
  10. F. Valtot, J. Kopel, E. Petit, F. Moulin, and J. Haut, “Treatment of refractory glaucoma with high-intensity focused ultrasound,” Journal Francais D Ophtalmologie, vol. 18, no. 1, pp. 3–12, 1995. View at: Google Scholar
  11. M. Lafond, F. Aptel, J. L. Mestas, and C. Lafon, “Ultrasound-mediated ocular delivery of therapeutic agents: a review,” Expert Opinion on Drug Delivery, vol. 14, no. 4, pp. 539–550, 2017. View at: Publisher Site | Google Scholar
  12. A. Razavi, D. Clement, A. R. Fowler et al., “Contribution of inertial cavitation in the enhancement of in vitro transscleral drug delivery,” Ultrasound in Medicine and Biology, vol. 40, no. 6, pp. 1216–1227, 2014. View at: Publisher Site | Google Scholar
  13. F. Aptel, T. Charrel, X. Palazzi, J. Y. Chapelon, P. Denis, and C. Lafon, “Histologic effects of a new device for high-intensity focused ultrasound cyclocoagulation,” Investigative Ophthalmology & Visual Science, vol. 51, no. 10, pp. 5092–5098, 2010. View at: Publisher Site | Google Scholar
  14. F. Aptel, T. Charre, C. Lafon et al., “Miniaturized high-intensity focused ultrasound device in patients with glaucoma: a clinical pilot study,” Investigative Ophthalmology & Visual Science, vol. 52, no. 12, pp. 8747–8753, 2011. View at: Publisher Site | Google Scholar
  15. Y. Li, J. Marie, B. Alfyan et al., “A novel model of persistent retinal neovascularization for the development of sustained anti-VEGF therapies,” Experimental Eye Research, vol. 174, pp. 98–106, 2018. View at: Publisher Site | Google Scholar
  16. R. B. Sampaio, R. J. Mendonca, A. R. Simioni et al., “Rabbit retinal neovascularization induced by latex angiogenic-derived fraction : an experimental model,” Current Eye Research, vol. 35, no. 1, pp. 56–62, 2010. View at: Publisher Site | Google Scholar
  17. L. M. Jiang, R. Chen, J. Xing et al., “Fabrication of a (K, Na)NbO3-based lead-free 1-3 piezocomposite for high-sensitivity ultrasonic transducers application,” Journal of Applied Physics, vol. 125, no. 21, p. 214501, 2019. View at: Publisher Site | Google Scholar
  18. C. M. Sayers and C. E. Tait, “Ultrasonic properties of transducer backings,” Ultrasonics, vol. 22, no. 2, pp. 57–60, 1984. View at: Publisher Site | Google Scholar
  19. R. Chen, L. Jiang, T. Zhang et al., “Eco-friendly highly sensitive transducers based on a new KNN-NTK-FM lead-free piezoelectric ceramic for high-frequency biomedical ultrasonic imaging applications,” IEEE Transactions on Biomedical Engineering, vol. 66, no. 6, pp. 1580–1587, 2019. View at: Publisher Site | Google Scholar
  20. X. J. Qian, T. Ma, M. Y. Yu, X. Y. Chen, K. K. Shung, and Q. F. Zhou, “Multi-functional ultrasonic micro-elastography imaging system,” Scientific Reports, vol. 7, no. 1, p. 1230, 2017. View at: Publisher Site | Google Scholar
  21. E. Y. Zernii, V. E. Baksheeva, E. N. Iomdina et al., “Rabbit models of ocular diseases : new relevance for classical approaches,” CNS & Neurological Disorders-Drug Targets (Formerly Current Drug Targets-CNS & Neurological Disorders), vol. 15, no. 3, pp. 267–291, 2016. View at: Publisher Site | Google Scholar
  22. L. Hyman, “Epidemiology of eye disease in the elderly,” Eye, vol. 1, no. 2, pp. 330–341, 1987. View at: Publisher Site | Google Scholar
  23. H. Ameri, G. J. Chader, J. Kim, S. R. Sadda, N. A. Rao, and M. S. Humayun, “The effects of intravitreous bevacizumab on retinal neovascular membrane and normal capillaries in rabbits,” IOVS, vol. 48, no. 12, pp. 5708–5715, 2007. View at: Google Scholar