Lu Zhang, Huijun Wang, Jianyi Liu, Shuang Chen, He Yang, Zewen Yang, Zhenxi Zhang, Hong Zhao, Li Yuan, Lifang Tian, Bo Zhong, Xiaolong Liu, "Scattering Inversion Study for Suspended Label-Free Lymphocytes with Complex Fine Structures", BME Frontiers, vol. 2022, Article ID 9867373, 14 pages, 2022. https://doi.org/10.34133/2022/9867373

Scattering Inversion Study for Suspended Label-Free Lymphocytes with Complex Fine Structures

Received29 May 2022
Accepted28 Sep 2022
Published08 Nov 2022

Abstract

Objective and Impact Statement. Distinguishing malignant lymphocytes from normal ones is vital in pathological examination. We proposed an inverse light scattering (ILS) method for label-free suspended lymphocytes with complex fine structures to identify their volumes for pathological state. Introduction. Light scattering as cell’s “fingerprint” provides valuable morphology information closely related to its biophysical states. However, the detail relationships between the morphology with complex fine structures and its scattering characters are not fully understood. Methods. To quantitatively inverse the volumes of membrane and nucleus as the main scatterers, clinical lymphocyte morphologies were modeled combining the Gaussian random sphere geometry algorithm by 750 reconstructed results after confocal scanning, which allowed the accurate simulation to solve ILS problem. For complex fine structures, the specificity for ILS study was firstly discussed (to our knowledge) considering the differences of not only surface roughness, posture, but also the ratio of nucleus to the cytoplasm and refractive index. Results. The volumes of membrane and nucleus were proved theoretically to have good linear relationship with the effective area and entropy of forward scattering images. Their specificity deviations were less than 3.5%. Then, our experimental results for microsphere and clinical leukocytes showed the Pearson product-moment correlation coefficients (PPMCC) of this linear relationship were up to 0.9830~0.9926. Conclusion. Our scattering inversion method could be effectively applied to identify suspended label-free lymphocytes without destructive sample pretreatments and complex experimental systems.

1. Introduction

Lymphocytes of human peripheral blood are an important indicator for detecting cellular immunity and humoral immunity, which is of great significance in the diagnosis of some diseases, such as malignant tumors [1], autoimmune and immune deficiency diseases [2], and blood diseases [3]. Labeling methods for lymphocyte organelles detection are widely applied for no matter traditional microscopic examination and flow cytometry [4, 5] even advanced super-resolution microscopy [6]. Unfortunately, the phototoxic and photobleaching problems in labeling detections make big challenges in maintaining the intrinsic biophysical states of cells. There are some cases that are not suitable for labeling technology. For example, it cannot be applied to the lipid [7]. Furthermore, cells growth states in the immunotherapy [810] should not be disturbed by any labeling processes. Because of cell’s intrinsic properties are closely related to its pathological process, the requirement that the intrinsic properties of cells can be detected without any labeling and other destructive treatments has been proposed. This problem has attracted great attention in the field of biological detections [1114]. Light scattering is closely related to cell’s biophysical states; therefore, it is called as cell’s “fingerprint” [15]. When a light beam illuminates on a label-free cell, its spatial scattered light carries its identity characteristics (volume, structures, components, et al.). It has been proved cell’s volume change has close relationship with many malignant diseases [1621]. For example, when patients suffer from the acute, chronic, and immature lymphocytic leukemia, the number of immature lymphocytes in blood increases with varying degrees [16, 17]. Compared with the average diameter 7.5 μm of mature lymphocytes [18], immature lymphocytes increase to about 10-16 μm. On the other hand, nucleus is another important indicator for cytopathologic diagnosis. It has been found that lymphocyte nuclei volumes of autism children decrease to 70% of healthy children [19], but for leukemia patients they become larger [20, 21]. Therefore, finding out the relationships between scattering features and cell’s morphology plays an important role for label-free biological detections, which is rooted in light scattering inversion (LSI) problem.

As early as 1970s, the relative volume distributions of biological particles with different diameters were estimated by one-dimensional (1D) scattering simulation method in which forward light scattering pattern (LSP) in small angle ranges was analyzed [22]. But at that time, how to detect LSP in experiments was a big challenge. LSP Fourier transform method was then put forward and experimentally verified by Semyanov et al. with the help of scanning flow cytometry [23]. In the meantime, the scattering phase function was studied to inverse biological particle sizes by a rotational confocal imaging setup [24]. Then, a further theoretical analysis in scattering inversion was done by Romanov and Yurkin [25]. However, these researches were mainly focused on 1D LSP of simple spherical particles. Su group tried to figure out the relationships of two-dimensional (2D) scattering images with biological particle sizes [2629]. Monte Carlo algorithm was designed to simulate the distributions of spherical organelles (such as mitochondrion) inside the membrane [28, 29]. The 2D scattering speckles and fringes of label-free spherical cells in a microchannel were imaged to inverse their internal structures [30, 31]. However, spheres and ellipsoids used to approximate cells’ biological structures were quite different from their real morphologies. Up to now, the detail relationships between cell’s real morphologies and its spatial scattering are not fully understood in LSI problem. The lack of accurate and reliable inversion principles makes scattering method hard to detect label-free clinical cells.

In this study, the real morphologies with complex fine internal and external structures of clinical lymphocytes in human peripheral blood were modeled instead of normally used spheres and ellipsoids. The relationships between their 2D scattering images and volumes of membrane and nucleus in LSI problem were studied, respectively. According to our research results, we found the area values in effective boundary of forward scattering image linearly increased with cell’s membrane volumes. Although nucleus volume was normally difficult to be detected without labeling methods, its scattering image entropy calculated by the gray level cooccurrence matrix algorithm (GLCM) was proved by us to have monotonic relationship with its volume. Furthermore, LSI specificity was firstly (to our knowledge) discussed in this paper to verify the inversed relationship between scattering features and cell’s biological structures. For complex fine structures, cell’s posture, refractive index, surface fold, and the ratio nucleus to the cytoplasm were discussed in detail. In order to verify the correctness of our scattering inversion principles, scattering experiments for both standard polystyrenes microspheres and clinical leukocytes from human peripheral blood were designed. The results showed the good linear and monotonic inversion relationships between scattering features and biological structures, which proved the good ability of our method in scattering inversion for suspended label-free lymphocytes with complex fine structures.

2. Materials and Methods

2.1. Materials
2.1.1. Standard Polystyrene Microspheres (SPM)

The commercial Duke polymer microspheres (Thermo Fisher Scientific Co. Ltd., USA) were used as the standard samples, which diameters, volumes, and other information are listed in Table 1.


NumberDiameter (μm)Volume (μm3)Original concentration (per ml)Std. dev (μm)C.V

133.510.0211.0%
266.280.0501.0%
3113.490.0601.0%
4178.210.0701.0%

Std. dev: standard deviation; C.V: coefficient of variance.
2.1.2. Lymphocytes in Human Peripheral Blood

Fresh human peripheral blood (2 ml) from healthy volunteer was collected in 5 ml EDTA vacutainer tube (Shijiazhuang Kangweishi Medical Equipment Co. Ltd., China). The blood was diluted by 2 ml phosphate buffer saline (PBS, pH 7.4) and then mixed with 4 ml lymphocyte separation medium (LSM, Tian Jing Haoyang Biological Manufacture Co. Ltd., China). The mixture was layered after centrifugation ( for 25 minutes) at room temperature. From the top to bottom of the layered mixture, there were plasma, lymphocytes, LSM, and erythrocytes. The layered lymphocytes were extracted by the pipette and washed with 5 ml PBS. After the second centrifugation ( for 5 minutes), lymphocytes gathered at the bottom were obtained by removing the supernatant fraction. Normally, the diameters of the obtained lymphocytes range from 4.8 to 12.0 μm [18] with the average value 7.5 μm.

2.2. Methods

Label-free lymphocyte scattering was studied to inverse the volumes of their membrane and nucleus, which is explained in Figure 1. As shown in Figures 1(a) and 1(b), morphological modeling and light scattering modeling should be carried out, respectively. Then, the area inside effective boundary of simulated forward scattering images and GLCM’s entropy of the region of interest (ROI) were calculated as two important parameters to inverse theoretically the volumes of label-free cells’ membranes and nuclei, as shown in Figures 1(c) and 1(d). In order to verify the theoretical inversion principles, forward scattering experiments for SPM and lymphocytes were designed, and scattering images were recorded as Figure 1(e). The inversion principles in Figure 1(f) were obtained. Finally, the experimental results were compared with theoretical inversion principles to assess its reliability and correctness.

2.2.1. Morphology Modeling

In order to study the inversion principles of suspended label-free lymphocytes with complex fine structures, 3D morphology modeling was very necessary. Different from traditional modeling methods which applied spheres and ellipsoids to represent the internal and external cellular structures, in this study clinical lymphocyte morphologies in suspension were detected layer by layer firstly by fluorescence confocal microscope (FCM). Their 3D morphologies of membranes and nuclei were reconstructed by our methods published before [32, 33]. The membrane and nucleus morphologies of clinical lymphocytes then were obtained as the basic morphology models. To generate different fine structures, the Gaussian random sphere geometry algorithm was introduced herein to make basic morphologies have complex fine structures, such as the fold and roughness on surface. The Gaussian random sphere geometry algorithm could generate a statistically related class for irregularly shaped objects, which had been used to model small objects, such as ice crystals [34], sand particles [35], and comets [36]. To our best knowledge, it seldomly applied in biological objects.

Complex fine structures generated by Gaussian random sphere geometry algorithm can be described as covariance function with logarithmic radius. A basic lymphocyte model is divided into discrete points. In spherical coordinate, their radii are modeled by Gaussian random sphere geometry algorithm, by which the radius of each discrete point follows the multivariate distribution.

in spherical coordinate has the relationship with its logarithmic radius as follows: wherein is their mean radius. is the radius relative standard deviation. The logarithmic radii for all discrete points can be expressed as . Corresponding to the given spherical coordinate, satisfies the multivariate normal statistical distribution: where is defined as the covariance matrix of logarithmic radii where the superscript represents the logarithmic radii. And can be expanded into the following equation: wherein is the angular distance between two directions and , and is the correlation function for logarithmic radius. can be expanded with a Legendre series . The Legendre series has been truncated from the lowest order to the highest order . Considering the modeling accuracy and simulation time, and are set to 2 and 50. The superscript is the power law coefficient. According to Equations (1)–(3), , , and are important to determine the radii for complex fine structures. controls the fluctuation degree, and controls the number of fold fluctuations in unit solid angle. The parameter () is selected to describe the finally generated fine structure features on 3D basic morphology models.

2.2.2. Scattering Modeling

FDTD algorithm is based on the finite differences instead of differentials to solve Maxwell’s equations, which was selected in this study to simulate the scattering distributions of lymphocytes in space. According to our former researches [37], in scattering modeling, the illuminating incident light was set with the wavelength along the forward direction of axis as shown in Figure 2. The amplitude normalization was set to 1. Cell nucleus and membrane were located in the position of coordinate origin to form the standard nucleus-membrane double-layer model. Small organelles (mitochondrion) were located based on the reconstructed results by above morphology modeling method. As shown in Figure 2(a), the far field scattering receiver was built as a cube. The scattering receiver plane CDHG was the forward scattering plane which was used in our scattering inversion, and ABCD was the sideward scattering plane. The perfectly matched layer boundary condition (PML) was adopted and in order to guarantee the correctness and accuracy. The simulating calculation step size was set as . Figure 2(b) presents the forward and sideward scattering modeling results for lymphocyte models with different structures.

In order to verify the correctness of our programs and parameter setting, the LSPs of microspheres with radius of 0.2 μm, 0.5 μm, and 1.0 μm calculated by our FDTD were compared with the results of Mie theory (classical algorithm for solving spherical media scattering), which is shown in Figure 3. It could be seen that LSPs of our FDTD method were in good agreement with classical Mie simulation. The relative errors for different microspheres were all within 0.92%, which confirmed the reliability and accuracy of our FDTD programs.

2.2.3. Experimental System

Experimental system was consisted of three parts: the illumination section by optical fiber laser, the scattering imaging section by microscope, and the sample fixing section by self-designed retainer, as shown in Figure 4(a). The single-mode optical fiber (Changfei Optical Fiber and Cable Co. Ltd., , China) was selected to transmit incident light onto a single lymphocyte. One end (A end in Figure 4(a)) of the single-mode optical fiber coupled with the fiber laser which wavelength and power are 650 nm and 2 mW, respectively (Beijing Xinglin Ruiguang Technology Co. Ltd., China). After polishing, the other end (B end) of the fiber was fixed by the sample retainer and faced perpendicularly to the bottom of glass slide. A drop of lymphocyte solution was dripped onto the grass slide which was fixed on the top of a 3D displacement table. In our experiments, the distance from B end to the bottom of the glass slide was about 12 mm. Glass slide, the retainer, single-mode optical fiber, and the 3D displacement table were all put on microscope stage. During experiments, a single lymphocyte was firstly observed clearly in bright field without oil immersion. Then, turn on the laser and adjust B end carefully to illuminate only a single lymphocyte. After turning off the bright field, forward scattering images (size: pixel) were collected on defocus plane (defocus distance about 30 micrometers). By image processing, the scattering features for scattering inversion principles could finally be obtained.

2.2.4. Image Processing

The effective area in forward scattering image was used for scattering inversion of membrane volume. The pixels whose gray values were about 57% of the maximum were defined as effective boundary. The area inside the effective boundary was the effective area. For scattering inversion of nucleus volume, GLCM algorithm was put forward to calculate the eigenvalues of ROI. GLCM is the matrix [38, 39] which row and column numbers are equal to the number of image grayscale levels. Its matrix element represents the occurrence probability of pixel pairs with gray values and . For GLCM, there are three key parameters , , and . is the original image gray level. is the distance between pixel pairs when GLCM is generated, which equals to 1, 2, 3, 4, 5…. And indicates the positional relationship of pixel pairs, usually selected as 0°, 45°, 90°, and 135°. The grayscale of our scattering image is . In order to reduce calculation time, the grayscale of scattering image should be compressed but without losing many important details. The gray level can be compressed to 16, 32, 64, and 128. Their computational times of extracting eigenvalues are shown in Table 2. Considering the computational efficiency, the gray level was finally determined to be 16, and equaled to 1. To ensure the changes of image texture in all directions were included, the final eigenvalues were the average values in the four directions of 0°, 45°, 90°, and 135°. In our experiments, four commonly used eigenvalues in describing texture features were selected:


Gray level163264128

Time (s)5.2619.0174.36289.00

3. Results and Discussion

3.1. 3D Modeling Results for Lymphocytes with Complex Fine Structures

750 lymphocytes in suspension from fresh human peripheral blood were detected by fluorescence confocal microscope. 3D lymphocyte models with five groups of different complex fine structures were constructed, which included the models with different membrane volumes but without nucleus (as shown in Figure 5(a)), the models with both different volumes and morphologies of membrane but without nucleus (as shown in Figure 5(b)), the models with different roughness on membrane’s surface but without nucleus (as shown in Figure 5(c)), the models with different nucleus volumes but with the same membranes (as shown in Figure 5(d)), and the models with both different volumes and morphologies but with different membrane volumes (as shown in Figure 5(e)).

3.2. Simulated Scattering Inversion Principles
3.2.1. Membranes with Different Volumes

Lymphocyte membrane models with the same surface folds but without nucleus were built, and their volumes were 157.0 μm3, 238.5 μm3, 344.1 μm3, 477.2 μm3, 640.7 μm3, and 837.8 μm3, respectively, as shown in Figure 6(a). Their membrane refractive index was set as 1.3675, and the membrane roughness was 0.48. The simulated forward scattering results are shown in Figure 6(b). The effective boundaries in Figure 6(b) are extracted and shown in Figure 6(c). After calculating effective area, a good linear relationship with the increase of lymphocyte membrane volumes was found (seeing Figure 6(d)). In order to evaluate its linear correlation, the commonly used parameter PPMCC was selected. For perfect negative and positive correlation, PPMCC equals to -1 and +1, but for no correlation, it equals to 0. In Figure 6(d), PPMCC for our scattering inversion was up to 0.9979, which meant lymphocyte membrane volume and its effective area in forward scattering image had a good linear inversed principle.

To evaluate the specificity of this linear relationship, some other biometric differences for lymphocyte complex fine structures were considered. Because human lymphocytes have a large ratio of nucleus to the cytoplasm, the influence of the nuclei on this linear inversion principle should be firstly discussed. A control group with five lymphocyte models had the same membrane volume (524.6 μm3) and refractive index (1.3675), but different nucleus volumes were set up in Figure 7(a). Their fold roughness was kept unchanged. The simulated forward scattering images and processed effective boundaries are shown in Figures 7(b) and 7(c). The effective area is then plotted in Figure 7(d). It could be known that the relationship between effective area and nucleus volumes was almost unchanged as displaying in Figure 7(d) like a horizontal line. And for different nucleus volumes, the effective area changed only less than 3.1%. It meant that the effective area had good specificity for membrane volume in scattering inversion.

Since suspended living lymphocytes always change their postures, the influence of postures needs to be considered. A group of models with randomly generated morphologies but controlled volumes in Figure 8 were discussed. It could be seen in Figure 8(d) that their relationship between membrane volume and its effective area in forward scattering image was still linear, and PPMCC for Figure 8(d) was calculated to be 0.9911. So, it proved the inversion principle was not affected by different postures, and the effective area was well specific to membrane volume.

On the other hand, the influence of membrane surface fold on the specificity of this scattering inversion principle was considered. Lymphocyte models with different membrane roughness are set up in Figure 9(a). Their simulated scattering images and effective area results are given in Figures 9(b) and 9(c). The deviation is defined as was selected to quantitatively evaluate the influence on the specificity. in this equation stood for the effective area, and was the membrane volume. According to the results in Figure 9(d), the influence of surface fold was not obvious, and the deviation between the maximum and the minimum was less than 3.5%.

Therefore, the conclusion could be drawn that the effective area of forward scattering image was the specific parameter to identify membrane volumes of label-free lymphocytes in suspension.

3.2.2. Nucleus with Different Volumes

To discuss the inversion principle for nucleus from scattering images, five lymphocyte models are set up in Figure 10(a) with different nucleus volumes but the same membrane volume (524.6 μm3), refractive index (1.3675 for membrane and 1.4000 for nucleus), and surface folds (0.48 for membrane and 0.41 for nucleus). Their nucleus volumes were 13.7 μm3, 33.2 μm3, 65.4 μm3, 113.5 μm3, and 180.5 μm3, respectively. ROI (in the red dot squares of Figure 10(b)) was an inner knot square region inside the effective boundary. Their zoomed-in images of ROI are shown in Figure 10(c). The typical four GLCM characters of ROI, angular second moment (ASM), contrast, inverse difference moment (IDM), and entropy, are given out in Figure 10(d). As mentioned in Section 3.2.1, although the effective area of forward scattering image was not sensitive to nucleus volumes, the entropy of ROI in Figure 10(d) increased with nucleus volume, and they had a good linear relationship.

In order to evaluate the specificity of nucleus volume and the entropy, lymphocyte models with both different volumes and various nucleus morphologies are set up in Figure 11(a). Their extracted ROIs and zoomed-in results are given out in Figure 11(b) with the red dot squares and Figure 11(c). Because the membrane volumes in this group were different, their ROI pixel sizes varied from , , , , and , respectively. And their GLCM entropy results are calculated and then plotted in Figure 11(d). Although in this group of lymphocyte models the membrane volumes changed, the linear monotonic relationship between nucleus volume and ROI entropy was still kept. PPMCC for their linear relationship was up to 0.9452. Therefore, ROI entropy could be determined as the specific parameter to inverse nucleus volumes of label-free lymphocytes in suspension.

3.2.3. Organelles with Different Refractive Indexes

Leukocytes are composed of various organelles. Besides their membranes and nuclei, organelles with different refractive indexes should also be considered. A group of organelles with different refractive indexes and volumes is modeled in Figure 12(a). Their refractive indexes were set as 1.3675, 1.3775, 1.3875, and 1.400, respectively, and the surface roughness () and posture were the same. After FDTD simulation and image processing, their forward scattering images and effective boundary are obtained in Figures 12(b) and 12(c), and effective area values are plotted in Figure 12(d). According to the clinical statistics information, the models #1 to #3 with diameter 5~8 μm had the similar size to human small lymphocytes, and the models #4 to #6 as medium to large lymphocytes. In Figure 12(d), it could be known that the refractive index had less influence on effective area of the models #1 to #3 in the blue box with small volumes. Such as the model #1 (membrane volume: ), for different refractive indexes in Figure 12(d), its effective areas were almost the same. In fact, except membrane and nucleus, no organelles could have such large volume because their diameters were only several hundreds of nanometers. Therefore, the influence of refractive index in scattering inversion could be ignored for a single organelle. Even when the organelles like mitochondria gathered as a group due to their different distributions, the whole volume was hard to be equal to that of model #1. Another case was the aggregation of several nuclei. Under this circumstance, their whole volume became large, but it was still less than that of model #6 which influence of refractive index on effective area could be seen as shown in the red box of Figure 12(d). The volume of the model #4 to #6 was similar to that of medium and large human lymphocytes. In order to assess the influence of refractive indexes on effective area for the model #4 to #6, their deviations were calculated. As a result, the maximum of four different refractive indexes were still less than 5.0%. It showed that on the one hand, effective area was not sensitive to refractive index for both larger membrane, nucleus, and small organelles, and on the other hand, its specificity for membrane volume was proved indirectly.

In fact, the components of human lymphocyte organelles are mainly lipids and proteins, whose refractive indices are about 1.36~1.37, such as the black line () and the red line () in Figure 12(d). Compared their area values on the red and black lines, their maximum for the model #1 to #6 was also less than 3.4%. So, it could be known that the difference on refractive index had little effect on scattering inversion between organelle volume and effective area.

3.3. Experimental Verification for Inversion Principle

Two kinds of samples were selected to verify the scattering inversion principles. The first one was the SPM in suspension with known morphology and diameter, which could quantitatively assess the precision of our scattering inversion principles. And the second one was the clinical label-free lymphocyte in vivo from human peripheral blood by which the effectiveness of scattering inversion principles in practical applications could be verified.

Suspended SPMs in Figure 13(a) (microscope observed) with diameters of 4.000 μm, 5.021 μm, 6.007 μm, and 6.982 μm (corresponding volumes: 33.5 μm3, 66.2 μm3, 113.1 μm3, and 178.2 μm3) were adopted for scattering imaging experiments, and their forward scattering images (in Figure 13(b)) were captured five times. The average effective area results (in Figure 13(c)) were calculated. The relationship between SPM volume and average effective area is shown in Figure 13(d). Eight sphere-like lymphocytes from human peripheral blood were also imaged. Their morphologies under microscopic observation are shown in Figure 14(a). Since the volumes of lymphocytes were unknown, so the cell was thought as an ellipsoid formed by the rotation of the focal plane image (Figure 14(a)) about axis , where and were the major and minor axis lengths on the focusing plane, respectively. As a result, the calculation equation of elliptical volume was applied to approximate that of lymphocytes. The edge enhancement process is done and shown in Figure 14(d) due to the low contrast between effective boundary and the background. Then, effective area values in Figure 14(e) are calculated and plotted in Figure 14(f). It could be seen from Figure 13(d) and Figure 14(f) that there were good linear relationships between the volume and effective area for both SPMs and clinical lymphocytes. The PPMCC was up to 0.9830 for SPM and 0.9926 for lymphocytes. Therefore, the scattering inversion principles concluded in Section 3.2 for suspended label-free lymphocytes with complex fine structures were well approved.

4. Conclusions

In this study, we proposed a method for inversing membrane and nucleus volume of label-free lymphocytes in suspension. It had been proved that the effective area in forward scattering image is closely related to its membrane volume, which was not sensitive to the surface roughness, the posture, and the ratio of nucleus to the cytoplasm and refractive index. And the entropy of ROI in forward scattering image had the ability to inverse nucleus volume, which was normally difficult to be detected without labeling method. The scattering image inversion principles in this study could be easily applied to identify suspended label-free lymphocytes in human peripheral blood, and the viability and intrinsic biological characters of cells could be well preserved. Our method also could be extended to other kinds of cells. On the other hand, our experimental system simply consisted of an ordinary microscope and a fiber laser. Compared with the labeling detection methods, it reduced the complexity of experimental system and sample pretreatments. In the further study, this method could be combined with machine learning algorithms to directly identify pathological lymphocytes from normal ones based on the volume differences of membrane and nucleus. And different kinds of microfluidic chips can be easily integrated into our experimental system, by which both the identification and isolation of label-free cells will become rapid and simple.

Data Availability

The data of scattering images to support the results of this study are available on request from the corresponding authors.

Conflicts of Interest

The authors declare that there is no conflict of interest regarding the publication of this article.

Authors’ Contributions

Lu Zhang designed experiments for these studies and wrote this manuscript. Huijun Wang conducted experiments, analyzed data, and drew the figures and tables. Jianyi Liu wrote and edited this manuscript. Shuang Chen conducted experiments and programmed the feature extraction algorithm. He Yang built the optical measurement platform. Zewen Yang assisted with simulations and experiments. Zhenxi Zhang provided guidance and support of optical measurement of biological cells. Hong Zhao provided guidance and support of optical scattering measurement technology. Li Yuan explained the data of simulation and assisted in the design of the experiment. Lifang Tian wrote the introduction to the medical background related to cytopathies and provided guidance on relevant medical knowledge. Bo Zhong conducted experiments on lymphocyte isolation and extraction. Xiaolong Liu conducted the design of the study and wrote and edited this manuscript.

Acknowledgments

The authors would like to thank the Xi’an Jiaotong University, School of Mechanical Engineering, State Key Laboratory for Manufacturing Systems Engineering, the Institute of Artificial Intelligence and Robotics, the Key Laboratory of Biomedical Information Engineering of Ministry of Education, the First Affiliated Hospital, the Second Affiliated Hospital, and the Mengchao Hepatobiliary Hospital of Fujian Medical University, The United Innovation of Mengchao Hepatobiliary Technology Key Laboratory of Fujian Province for their support of this work. This work was supported by the National Natural Science Foundation of China (Grant Number: 61875160).

References

  1. K. Shanta, K. Nakayama, M. Ishikawa, T. Ishibashi, and S. Kyo, “Prognostic value of peripheral blood lymphocyte telomere length in gynecologic malignant tumors,” Cancers, vol. 12, no. 6, 2020. View at: Publisher Site | Google Scholar
  2. S. G. Tangye, S. J. Pelham, E. K. Deenick, and C. S. Ma, “Cytokine-mediated regulation of human lymphocyte development and function: insights from primary immunodeficiencies,” Journal of Immunology, vol. 199, no. 6, pp. 1949–1958, 2017. View at: Publisher Site | Google Scholar
  3. U. Gerdemann, A. S. Christin, C. R. Cruz, C. A. Ramos, and A. M. Leen, “Broad spectrum tumor antigen-specific cytotoxic T lymphocytes (CTL) for therapy of hematological malignancy,” Blood, vol. 114, no. 22, pp. 4083–4083, 2009. View at: Publisher Site | Google Scholar
  4. B. J. Bain, “Diagnosis from the blood smear,” New England Journal of Medicine, vol. 353, no. 5, pp. 498–507, 2005. View at: Publisher Site | Google Scholar
  5. Z. Darzynkiewicz, E. Bedner, and P. Smolewski, “Flow cytometry in analysis of cell cycle and apoptosis,” Seminars in Hematology, vol. 38, no. 2, pp. 179–193, 2001. View at: Publisher Site | Google Scholar
  6. Z. Tian, Y. Wei, Y. Yu, L. Fu, and Z. L. Hu, “Blood cell analysis: from traditional methods to super-resolution microscopy,” Photonics, vol. 9, no. 4, pp. 261–261, 2022. View at: Publisher Site | Google Scholar
  7. D. Hua, S. Qi, H. Li, Z. Zhang, and L. Fu, “Monitoring the process of pulmonary melanoma metastasis using large area and label-free nonlinear optical microscopy,” Journal of Biomedical Optics, vol. 17, no. 6, article 066002, 2012. View at: Publisher Site | Google Scholar
  8. U. Schnell, F. Dijk, K. A. Sjollema, and B. N. G. Giepmans, “Immunolabeling artifacts and the need for live-cell imaging,” Nature Methods, vol. 9, no. 2, pp. 152–158, 2012. View at: Publisher Site | Google Scholar
  9. C. Snow, “Flow cytometer electronics,” Cytometry Part A, 2010, vol. 57A, no. 2, pp. 63–69, 2004. View at: Publisher Site | Google Scholar
  10. N. Nitta, T. Iino, A. Isozaki et al., “Raman image-activated cell sorting,” Nature Communications, vol. 11, no. 1, p. 3452, 2020. View at: Publisher Site | Google Scholar
  11. C. Liu, C. Capjack, and W. Rozmus, “3-D simulation of light scattering from biological cells and cell differentiation,” Journal of Biomedical Optics, vol. 10, no. 1, article 014007, 2005. View at: Publisher Site | Google Scholar
  12. J. Wilson, C. Bigelow, D. J. Calkins, and T. H. Foster, “Light scattering from intact cells reports oxidative-stress-induced mitochondrial swelling,” Biophysical Journal, vol. 88, no. 4, pp. 2929–2938, 2005. View at: Publisher Site | Google Scholar
  13. L. Xie, Q. Liu, C. S. Shao, and X. T. Su, “Automatic classification of acute and chronic myeloid leukemic cells with wide-angle label-free static cytometry,” Optics Express, vol. 25, no. 23, pp. 29365–29373, 2017. View at: Publisher Site | Google Scholar
  14. N. Nitta, T. Sugimura, A. Isozaki, H. Mikami, and K. Goda, “Intelligent image-activated cell sorting,” Cell, vol. 175, no. 1, pp. 266–276.e13, 2018. View at: Publisher Site | Google Scholar
  15. V. P. Maltsev, “Scanning flow cytometry for individual particle analysis,” Review of Scientific Instruments, vol. 71, no. 1, pp. 243–255, 2000. View at: Publisher Site | Google Scholar
  16. W. Xu, J. Y. Li, J. F. Zhang et al., “Clinical analysis of chronic lymphocytic leukemia with naive lymphocytosis,” Chinese Journal of Hematology, vol. 28, no. 11, pp. 780–782, 2007. View at: Publisher Site | Google Scholar
  17. C. E. Bueso-Ramos, A. Ferrajoli, L. J. Medeiros, M. J. Keating, and Z. Estrov, “Hematological malignancies,” Hematology, vol. 9, no. 4, pp. 279–286, 2004. View at: Publisher Site | Google Scholar
  18. A. Zharinov, P. Tarasov, A. Shvalov, K. Semyanov, D. R. van Bockstaele, and V. Maltsev, “A study of light scattering of mononuclear blood cells with scanning flow cytometry,” Journal of Quantitative Spectroscopy and Radiative Transfer, vol. 102, no. 1, pp. 121–128, 2006. View at: Publisher Site | Google Scholar
  19. N. E. Abdelmegui and S. G. Kourtian, “Subcellular disorders associated with autism peripheral blood,” Asian Journal of Biological Sciences, vol. 5, no. 5, pp. 221–239, 2012. View at: Publisher Site | Google Scholar
  20. D. Zink, A. H. Fische, and J. A. Nickerson, “Nuclear structure in cancer cells,” Nature Reviews Cancer, vol. 4, no. 9, pp. 677–687, 2004. View at: Publisher Site | Google Scholar
  21. H. Wang, F. Y. Ren, W. H. Peng, Y. T. Lan, and J. B. Wang, “Establishment of reference intervals of the white blood cell VCS parameters and their application in sepsis and leukemia,” Laboratory Medicine, vol. 27, no. 7, pp. 534–539, 2012. View at: Publisher Site | Google Scholar
  22. P. F. Mullaney and P. N. Dean, “The small-angle light scattering of biological cells: theoretical considerations,” Biophysical Journal, vol. 10, no. 8, pp. 764–772, 1970. View at: Publisher Site | Google Scholar
  23. K. A. Semyanov, P. A. Tarasov, A. E. Zharinov, A. V. Chernyshev, and V. P. Maltsev, “Single-particle sizing from light scattering by spectral decomposition,” Applied Optics, vol. 43, no. 26, pp. 5110–5115, 2004. View at: Publisher Site | Google Scholar
  24. J. L. Castagner and I. J. Bigio, “Particle sizing with a fast polar nephelometer,” Applied Optics, vol. 46, no. 4, pp. 527–532, 2007. View at: Publisher Site | Google Scholar
  25. A. V. Romanov and M. A. Yurkin, “Rigorous analysis of the spectral sizing of single particles based on light scattering patterns,” Optics & Laser Technology, vol. 151, article 108047, 2022. View at: Publisher Site | Google Scholar
  26. X. T. Su, C. Capjack, W. Rozmus, and C. Backhouse, “2D light scattering patterns of mitochondria in single cells,” Optics. Express, vol. 15, no. 17, pp. 10562–10575, 2007. View at: Publisher Site | Google Scholar
  27. X. T. Su, K. Singh, C. Capjack, J. Petracek, C. Backhouse, and W. Rozmus, “Measurements of light scattering in an integrated microfluidic waveguide cytometer,” Journal of Biomedical Optics, vol. 13, no. 2, article 024024, 2008. View at: Publisher Site | Google Scholar
  28. X. T. Su, K. Singh, W. Rozmus, C. Backhouse, and C. Capjack, “Light scattering characterization of mitochondrial aggregation in single cells,” Optics Express, vol. 17, no. 16, pp. 13381–13388, 2009. View at: Publisher Site | Google Scholar
  29. X. T. Su, Y. Qiu, L. Marquez-Curtis et al., “Label-free and noninvasive optical detection of the distribution of nanometer-size mitochondria in single cells,” Journal of Biomedical Optics, vol. 16, no. 6, article 067003, 2011. View at: Publisher Site | Google Scholar
  30. M. Lin, X. Qiao, Q. Liu, C. S. Shao, and X. T. Su, “Light-sheet-based 2D light scattering cytometry for label-free characterization of senescent cells,” Biomedical Optics Express, vol. 7, no. 12, pp. 5170–5181, 2016. View at: Publisher Site | Google Scholar
  31. M. Lin, L. Qiao, L. Chao, Q. Xu, and X. T. Su, “Label-free light-sheet microfluidic cytometry for the automatic identification of senescent cells,” Biomedical Optics Express, vol. 9, no. 4, pp. 1692–1703, 2018. View at: Publisher Site | Google Scholar
  32. L. Zhang, Y. Xie, Y. Z. Tu et al., “Clinical lymphocytes construction for light scattering inversion study: a three-dimensional morphology constructed method from defective confocal images,” Journal of Biomedical Optics, vol. 23, no. 8, 2018. View at: Publisher Site | Google Scholar
  33. Z. W. Yang, L. Zhang, N. Lv et al., “Accurate 3D morphological computational model reconstruction of suspended cells imaged through stratified media by the precise depth-varying point spread function method,” Optics Express, vol. 30, no. 15, pp. 27539–27559, 2022. View at: Publisher Site | Google Scholar
  34. T. Nousiainen and G. M. Mcfarquhar, “Light scattering by quasi-spherical ice crystals,” Journal of the Atmospheric Sciences, vol. 61, no. 18, pp. 2229–2248, 2004. View at: Publisher Site | Google Scholar
  35. T. Nousiainen, “Scattering of light by large Saharan dust particles in a modified ray optics approximation,” Journal of Geophysical Research Atmospheres, vol. 108, no. D1, 2003. View at: Publisher Site | Google Scholar
  36. K. Muinonen, “Introducing the Gaussian shape hypothesis for asteroids and comets,” Astronomy and Astrophysics, vol. 332, pp. 1087–1098, 1998. View at: Google Scholar
  37. L. Zhang, Y. W. Qin, K. X. Li et al., “Light scattering properties in spatial planes for label free cells with different internal structures,” Optical and Quantum Electrics, vol. 47, no. 5, pp. 1005–1025, 2015. View at: Publisher Site | Google Scholar
  38. R. M. Haralick, K. Shanmugam, and I. Dinstein, “Textural features for image classification,” Studies in Media and Communication, vol. SMC-3, no. 6, pp. 610–621, 1973. View at: Publisher Site | Google Scholar
  39. V. V. Starovoitov, S. Y. Jeong, and R. H. Park, “Texture periodicity detection: features, properties, and comparisons,” IEEE Transactions on Systems Man & Cybernetics Part A, vol. 28, no. 6, pp. 839–849, 1998. View at: Publisher Site | Google Scholar