BME Frontiers / 2022 / Article

Research Article | Open Access

Volume 2022 |Article ID 9891510 | https://doi.org/10.34133/2022/9891510

Qiangzhou Rong, Youngseop Lee, Yuqi Tang, Tri Vu, Carlos Taboada, Wenhan Zheng, Jun Xia, David A. Czaplewski, Hao F. Zhang, Cheng Sun, Junjie Yao, "High-Frequency 3D Photoacoustic Computed Tomography Using an Optical Microring Resonator", BME Frontiers, vol. 2022, Article ID 9891510, 8 pages, 2022. https://doi.org/10.34133/2022/9891510

High-Frequency 3D Photoacoustic Computed Tomography Using an Optical Microring Resonator

Received15 Nov 2021
Accepted25 May 2022
Published01 Aug 2022

Abstract

3D photoacoustic computed tomography (3D-PACT) has made great advances in volumetric imaging of biological tissues, with high spatial-temporal resolutions and large penetration depth. The development of 3D-PACT requires high-performance acoustic sensors with a small size, large detection bandwidth, and high sensitivity. In this work, we present a new high-frequency 3D-PACT system that uses a microring resonator (MRR) as the acoustic sensor. The MRR sensor has a size of 80 μm in diameter and was fabricated using the nanoimprint lithography technology. Using the MRR sensor, we have developed a transmission-mode 3D-PACT system that has achieved a detection bandwidth of ~23 MHz, an imaging depth of ~8 mm, a lateral resolution of 114 μm, and an axial resolution of 57 μm. We have demonstrated the 3D PACT’s performance on in vitro phantoms, ex vivo mouse brain, and in vivo mouse ear and tadpole. The MRR-based 3D-PACT system can be a promising tool for structural, functional, and molecular imaging of biological tissues at depths.

1. Introduction

Photoacoustic computed tomography (PACT), also referred to as optoacoustic tomography (OAT), is a hybrid biomedical imaging modality that combines optical excitation and ultrasound detection [15]. In PACT, three-dimensional (3D) images can be reconstructed using a matrix array of ultrasound sensors without mechanical scanning [46]. These ultrasound sensors have been used in preclinical applications with high-frame rate and high resolution. The matrix array transducer is often made from the piezoelectric materials [79]. Previous research has shown that, although the matrix array transducer can achieve fast data acquisition, it has some limitations for 3D-PACT applications. The detection sensitivity, quantified as the noise equivalent pressure (NEP), is inversely proportional to the sensor size [10]. The small sensor size, usually comparable to the acoustic wavelength, results in low detection sensitivity. Piezoelectric ultrasound sensors usually have limited detection frequency bandwidth and acceptance angle [1113]. The limited bandwidth eventually leads to relatively poor spatial resolution and increased reconstruction artifacts. To improve the detection bandwidth, multiple piezoelectric ultrasound sensors with different frequency bands can be integrated, which, however, increases the system complexity and reduces the imaging speed [13, 14]. Because of the omnidirectional nature of the PA waves, the limited acceptance angle of the piezoelectric ultrasound sensors also results in degraded image quality such as missing target features and limited-view artifacts [1517]. Moreover, to cover a large field of view (FOV), piezoelectric ultrasound sensors need to be placed at a large distance from the target, which leads to the loss of the high-frequency signals and further decreases the spatial resolution.

Optical ultrasound sensors have been explored to overcome the limited bandwidth and acceptance angle of the piezoelectric ultrasound sensors [1822]. A variety of optical ultrasound sensors with promising performance have been developed for a number of PACT applications, including high-finesse planar sensor [23], in-fiber Fabry-Pérot interferometer [19, 24], in-fiber laser sensor [25, 26], submicrometer sensors on a photonic chip [27, 28], and optomechanical ultrasound sensors [29]. In these optical ultrasound sensors, the PA signals are detected by measuring pressure-induced optical phase change over the optical path. The optical sensors usually have small sizes, high detection sensitivity, broad bandwidth, and high scalability for different imaging configurations [18]. Among them, the polymer-based microring resonator (MRR) offers unique advantage due to its miniaturized form factor and optical transparency [3033]. Its submicron thickness favorably scales its acoustomechanical resonance to the gigahertz range and, thus, allows uniform frequency response from DC to hundreds of megahertz. MRR has been used for PA microscopy of single cells [34] and longitudinal mouse brain imaging [35], with superficial imaging depth. However, the utility of MRR for deep PA imaging has not been demonstrated. In this paper, we have reported an MRR sensor being optimized for deep-tissue high-frequency 3D-PACT. Both the lateral and axial resolutions of PACT benefit from the wide detection bandwidth of the MRR sensor. The new MRR sensor is encapsulated by an acoustic impedance matched protection layer to improve its reliability and stability for in vivo applications. Using the new MRR sensor, we developed a transmission-mode 3D-PACT with a large FOV. We have characterized the MRR-based 3D-PACT system on phantoms and demonstrated its in vivo application on mice and tadpoles. We expect that the current work will pave the way for high-speed, high-resolution, deep-penetrating 3D-PACT using optical ultrasound sensors.

2. Methods and Materials

2.1. MRR-Based 3D-PACT System

The MRR-based 3D-PACT system is shown in Figure 1(a). A pulsed Nd: YAG laser (Q-smart 850, Quantel) was employed as the excitation light source at 532 nm, with a pulse repetition frequency (PRF) of 10 Hz. An optical diffuser (DG10-220-MD, Thorlabs) was used to expand and homogenize the light beam over the sample. The light beam had a diameter of ~1 cm on the sample surface. The MRR detector was mounted ~4 mm beneath the sample with water in between as the acoustic coupling medium. 3D PA imaging was achieved by raster scanning the sample, while the MRR sensor and the excitation light were kept stationary. The sample was mounted on a three-axis motorized scanning stage (L-509, PI) that can provide a scanning range over an FOV of 26 by 26 mm2. The scanning step size was 60 μm along the - and -axis. The total image acquisition time was jointly determined by the laser’s PRF, the scanning step size, and the FOV. A narrow-band tunable laser (TLB-6712, New Focus) was used as the interrogation light source for the MRR sensor, which provides an output wavelength range of 765–781 nm, a narrow linewidth of <200 kHz, and a wavelength tuning step size of 1.5 pm. The interrogation light at 772.4 nm was coupled into the MRR input through a single-mode fiber (SMF). The transmission light exiting the MRR output was collected by a multimode fiber (MMF) and detected by an avalanche photodetector (APD430A, Thorlabs). The photodetector signals were recorded by a data acquisition card (ATS9350, Alazar) with a sample rate of 500 MHz. We used a 3D delay-and-sum algorithm to reconstruct the volumetric PA images [3639]. The GPU-based (NVIDIA GeForce RTX) reconstruction took ~10 mins for an image with voxels. Hilbert transform was performed along the -axis of the reconstructed bipolar image to extract the signal envelope.

2.2. Fabrication of the MRR Sensor

The MRR sensor was fabricated using soft nanoimprint lithography (sNIL). The fabrication process mainly included three steps: (1) fabrication of a silicon (Si) master mold, (2) replication of a polydimethylsiloxane (PDMS) soft mold from the Si master mold, and (3) fabrication of the MRR sensor on a quartz substrate by sNIL [35]. The fabricated MRR sensor shown in Figures 1(b) and 1(c) consists of a microring waveguide (MRW) and a matching straight bus waveguide (BW). As the incident light from the BW evanescently coupled into the MRR, the resultant whispering gallery modes generate a strong optical resonance. The MRW fabricated from soft polymer then transduces the incoming acoustic pressure into the observable frequency shift of the optical resonance, which can be subsequently detected using a narrow-band laser source. The sensitivity of the MRR sensor is proportional to the quality factor (-factor) of the optical resonance. The -factor is defined by , where is the resonance wavelength and is the full width at half maximum (FWHM) of the measured resonance spectrum.

2.3. Optimization of the MRR Sensor Stability for In Vivo Imaging

It is critical to ensure the reliable operation of MRR against the physiological contaminants for in vivo PA imaging on small animals. Thus, as shown in Figures 1(d) and 1(e), a 6 μm thick low refractive index (RI) polymer (MY-131 MC, MY Polymers Ltd.) was spin-coated on the quartz substrate to fully encapsulate the MRR. MY-131 MC was chosen for three reasons: (1) its high biocompatibility; (2) its low RI of 1.312 that reduces the propagating loss in MRW and thus increases the -factor, compared to the PDMS cladding layer with a RI of 1.430 [35]; and (3) its low acoustic impedance that minimizes the ultrasound reflection loss. As shown in Figure 1(f), the fabricated MRR sensor was mounted on a transparent glass substrate for the convenience of handling. The simulated mode confinement within the waveguide is shown in the inset of Figure 1(g). The simulated resonance spectrum and its shift under ultrasound pressure are illustrated in Figure 1(g). Figure 1(h) shows the comparison of resonance spectra for MRRs with different cladding materials around the wavelength of 772.4 nm. Lorenz fitting indicates that the spectrum of MRR with the low RI cladding has a FWHM of 16.8 pm and -factor of , which is 1.4 times higher than that of the MRR with the PDMS cladding [35]. The MRR sensor supports different resonance modes at orthogonal polarization states: quasi-TE mode and quasi-TM mode. The experimentally measured resonance spectra of the MRR sensor at two states are shown in Figure 1(i). We chose the quasi-TM mode with a higher -factor in this study.

3. Results

3.1. MRR Sensor Characterization

The MRR sensor’s performance was characterized by using the 3D-PACT system. The spatial resolution was measured by imaging a carbon fiber with a diameter of 7 μm, which was thin enough to be considered as a line target. The carbon fiber was placed parallel to the -axis. The reconstructed PA image is shown in Figure 2(a). The signal profile was used to fit the line spread function (LSF) of the system. As shown in Figures 2(b) and 2(c), the FWHM of the LSF was measured as the spatial resolution along each axis. The lateral resolution, which was mainly determined by the central frequency and the effective detection aperture, was measured to be ~114 μm [24]. The axial resolution, which was majorly determined by the bandwidth of the MRR sensor, was measured to be ~57 μm. With the carbon fiber as a line target, the temporal response of the MRR sensor was recorded and is shown in Figure 2(d). The frequency response of the sensor was obtained from the temporal response, as shown in Figure 2(e). The −3 dB bandwidth of the frequency response profile was ~23 MHz. At a working distance of ~6.3 mm, the directivity of the sensor was measured by mapping the time-resolved PA signals at different lateral angles with respect to the sensor. As shown in Figure 2(f), an angular coverage of ~90° was achieved.

3.2. 3D-PACT on Phantoms

The MRR-based 3D-PACT was demonstrated first on several phantoms. First, four human hairs were placed parallel to each other at different depths in water (Figure 3(a) (i)). The 3D PA image was reconstructed as shown in Figure 3(a) (ii). The cross-sections of four hairs can be clearly resolved at different depths. The degraded lateral resolution at larger depth is mainly due to the reduced detection aperture. Second, we imaged two more phantoms in optically scattering medium: three crossed human hairs (Figure 3(b) (i)) and a black leaf skeleton (Figure 3(c) (i)). Both samples have rich directional features, mimicking the vascular structures in biological tissues. Both samples were embedded inside agar with 0.5% Intralipid. In Figure 3(b) (ii), all three hairs can be clearly imaged with a high contrast. In Figure 3(c) (ii), most of the leaf structural features were clearly imaged, especially the main skeletons. To further demonstrate 3D-PACT of arbitrary targets, we imaged four human hairs randomly distributed in volume, as well as a hair knot randomly oriented in scattering medium. Both phantoms can be clearly reconstructed with a penetration depth of up to ~8.5 mm, as shown in Figures 3(d) and 3(e).

3.3. 3D-PACT on Ex Vivo Tissue and In Vivo Animals

After validating the system performance on phantoms, we demonstrated 3D-PACT on ex vivo tissue and in vivo animal models. All research procedures on animals were approved by the Institutional Animal Care and Use Committee of Duke University (protocol No. A009-20-01). First, we perfused a mouse brain using a blood-gelatin mixture (50% whole bovine blood, 3.5% gelatin, and 46.5% water) and placed the perfused brain at 4°C for 30 mins to solidify the gelatin. The perfused brain was then embedded in agar with 0.5% Intralipid and imaged by the 3D-PACT system, as shown in Figure 4(a). The maximum intensity projection (MAP) images are shown in Figure 4(b), in which the major vasculature of the perfused brain could be resolved.

Second, as a proof of concept, we performed in vivo PA imaging of a mouse ear. The mouse was anesthetized with 1.0-1.5% isoflurane and kept warm at 37°C. We overlaid a thin layer of scattering medium on top of the mouse ear to homogenize the excitation light, as shown in Figure 4(c). The reconstructed MAP images are shown in Figure 4(d). The major vascular pairs in the ear are clearly visualized with high contrast. However, the microvessels cannot be resolved due to the limited resolutions. Third, a Xenopus laevis tadpole (Ward’s Science) with a body dimension of , was imaged in vivo, as shown in Figure 4(e). The tadpole was anesthetized by immersion in tricaine methanesulfonate solution (MS-222, Sigma-Aldrich) at a concentration of 0.35 g/L. The tadpole’s skin surface was exposed in water to enable normal breathing. Figure 4(f) shows the reconstructed MAP images of the tadpole, and we can clearly observe major blood vessels, eyes, brain, guts, and skin pigments.

4. Conclusion and Discussion

In conclusion, we have demonstrated a new implementation of 3D-PACT using a nanofabricated MRR sensor. The miniaturized MRR has a small sensing area of 80 μm in diameter, which allows MRR operating as a point-like detector. The MRR is fabricated on a quartz microscope coverslip, and thus, it is fully transparent and ideal for both transmission and reflective imaging modes. The nanofabrication process has been optimized to support strong optical resonance with a -factor of , leading to a high detection sensitivity with a noise equivalent pressure (NEP) of 81 Pa, despite such a small sensing area [30, 34]. In combination with its broad detection bandwidth (~23 MHz) and wide acceptance angle (90°), the 3D-PACT system can provide lateral and axial resolutions of ~114 μm and ~57 μm, respectively. A thin transparent protection layer was used to fully encapsulate the MRR sensor, for the benefit of its long-term stability in water and in vivo environment. Additionally, the low-cost sNIL method used to fabricate MRR allows such a highly sensitive ultrasound detector to be disposable for demanding in vivo imaging applications. Its miniaturized form factor is highly desirable for catheter and endoscope applications. All these properties, along with its high transparency for reflection mode imaging, make the MRR sensor an ideal candidate for 3D-PACT.

The performance of the reported MRR-based 3D-PACT system was validated using a set of experiments: human hairs and leaf skeleton in arbitrary geometries, an ex vivo perfused mouse brain, and an in vivo mouse ear and tadpole. The 3D reconstructed images of these phantoms were clearly resolved with high SNR and contrast. However, the current transmission mode implementation is limited for ex vivo samples and thin in vivo tissues. For example, it is challenging to perform transcranial brain imaging in the transmission mode, due to the strong ultrasound attenuation through the bone. One challenge for implementing the reflection mode is the interference between incident laser pulse and the MRR detection. Further optimizing the MRR sensor design to mitigate this issue would enable MRR-based 3D-PACT in the reflection mode. The imaging speed is currently limited by the low PRF of the excitation laser at 10 Hz. A laser with a higher PRF can improve the imaging speed. Furthermore, the 532 nm light was used for PA signal generation with limited penetration depth [40]. Imaging depth can be improved by using longer wavelength light, preferred in the near-infrared region. The conventional high-frequency piezoelectric ultrasound transducer can also achieve wide detection bandwidth. However, these high-frequency transducers typically cannot detect the low-frequency signals, while the MRR sensor has better low-frequency response (Figure 2(e)), which is critical for deep-tissue imaging. Moreover, with the same active sensor size, the MRR sensor has better detection sensitivity, more uniform frequency response, and larger acceptance angle than the piezo-based detector. Considering the MRR sensor as a point-like detector, the theoretical lateral and axial resolutions of our system are, respectively ~89 μm and~39 μm [41], which are close to the measured resolutions in clear medium. In practice, both the lateral and axial resolutions become worse with the increasing depth, mainly due to the decrease in the high-frequency acoustic signals and the detection aperture [42, 43]. Further miniaturization of MRR will allow it to function as a true point detector, which can improve the spatial resolution and angular response. All in all, we expect that the presented work can pave the way for volumetric PACT with high resolution, deep penetration, and high sensitivity and enable new clinical and preclinical biomedical applications.

Data Availability

The results used to support the findings of this study are included within the article. The raw datasets generated during the study are available for research purposes from the corresponding author on reasonable request.

Conflicts of Interest

The authors declare no conflicts of interest.

Authors’ Contributions

Qiangzhou Rong and Youngseop Lee contributed equally to this work.

Acknowledgments

This work was sponsored by American Heart Association Collaborative Sciences Award (18CSA34080277); the United States National Institutes of Health (NIH) grants R21EB027981, R21 EB027304, RF1 NS115581 (BRAIN Initiative), R01 NS111039, and R01 EB028143; and Chan Zuckerberg Initiative Grant (2020-226178), all to J. Yao. NIH grant P41GM135018 was awarded to H. Zhang and C. Sun. This work was performed, in part, at the Center for Nanoscale Materials, a U.S. Department of Energy Office of Science User Facility, and supported by the U.S. Department of Energy, Office of Science, under Contract No. DE-AC02-06CH11357.

References

  1. L. V. Wang, “Multiscale photoacoustic microscopy and computed tomography,” Nature Photonics, vol. 3, no. 9, pp. 503–509, 2009. View at: Publisher Site | Google Scholar
  2. L. V. Wang and S. Hu, “Photoacoustic tomography: in vivo imaging from organelles to organs,” Science, vol. 335, pp. 1462–1485, 2012. View at: Google Scholar
  3. L. Lin, P. Hu, J. Shi et al., “Single-breath-hold photoacoustic computed tomography of the breast,” Nature Communications, vol. 9, no. 1, p. 2352, 2018. View at: Publisher Site | Google Scholar
  4. L. Lin, X. Tong, P. Hu, M. Invernizzi, L. Lai, and L. V. Wang, “Photoacoustic computed tomography of breast cancer in response to neoadjuvant chemotherapy,” Advanced Science, vol. 8, no. 7, p. 2003396, 2021. View at: Publisher Site | Google Scholar
  5. A. Taruttis and V. Ntziachristos, “Advances in real-time multispectral optoacoustic imaging and its applications,” Nature Photonics, vol. 9, no. 4, pp. 219–227, 2015. View at: Publisher Site | Google Scholar
  6. L. Li, L. Zhu, R. Zhang et al., “Single-impulse panoramic photoacoustic computed tomography of small-animal whole-body dynamics at high spatiotemporal resolution,” Nature Biomedical Engineering, vol. 1, no. 5, 2017. View at: Publisher Site | Google Scholar
  7. S. Na, J. J. Russin, L. Lin et al., “Massively parallel functional photoacoustic computed tomography of the human brain,” Springer Nature, vol. 6, no. 5, pp. 584–592, 2022. View at: Publisher Site | Google Scholar
  8. L. V. Wang and J. Yao, “A practical guide to photoacoustic tomography in the life sciences,” Nature Methods, vol. 13, no. 8, pp. 627–638, 2016. View at: Publisher Site | Google Scholar
  9. Y. Zhang and L. Wang, “Video-rate ring-array ultrasound and photoacoustic tomography,” IEEE Transactions on Medical Imaging, vol. 39, no. 12, pp. 4369–4375, 2020. View at: Publisher Site | Google Scholar
  10. A. M. Winkler, K. Maslov, and L. V. Wang, “Noise-equivalent sensitivity of photoacoustics,” Journal of Biomedical Optics, vol. 18, no. 9, article 097003, 2013. View at: Publisher Site | Google Scholar
  11. A. Dangi, C. Y. Cheng, S. Agrawal et al., “A photoacoustic imaging device using piezoelectric micromachined ultrasound transducers (PMUTs),” Transactions on Ultrasonics Ferroelectrics, and Frequency Control, vol. 67, no. 4, pp. 801–809, 2020. View at: Publisher Site | Google Scholar
  12. G. Ku, X. Wang, G. Stoica, and L. V. Wang, “Multiple-bandwidth photoacoustic tomography,” Physics in Medicine & Biology, vol. 49, no. 7, pp. 1329–1338, 2004. View at: Publisher Site | Google Scholar
  13. C. Tian, M. Pei, K. Shen, S. Liu, Z. Hu, and T. Feng, “Impact of system factors on the performance of photoacoustic tomography scanners,” Physical Review Applied, vol. 13, no. 1, article 014001, 2020. View at: Publisher Site | Google Scholar
  14. J. Zhang, S. H. Pun, Y. Yu et al., “Development of a multi-band photoacoustic tomography imaging system based on a capacitive micromachined ultrasonic transducer array,” Applied Optics, vol. 56, no. 14, pp. 4012–4018, 2017. View at: Publisher Site | Google Scholar
  15. C. Tao and X. Liu, “Reconstruction of high quality photoacoustic tomography with a limited-view scanning,” Optics Express, vol. 18, no. 3, pp. 2760–2766, 2010. View at: Publisher Site | Google Scholar
  16. W. Shu, M. Ai, T. Salcudean, R. Rohling, P. Abolmaesumi, and S. Tang, “Broadening the detection view of 2D photoacoustic tomography using two linear array transducers,” Optics Express, vol. 24, no. 12, pp. 12755–12768, 2016. View at: Publisher Site | Google Scholar
  17. S. Acosta, “Solvability for photoacoustic imaging with idealized piezoelectric sensors,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 67, no. 11, pp. 2413–2422, 2020. View at: Publisher Site | Google Scholar
  18. G. Wissmeyer, M. A. Pleitez, A. Rosenthal, and V. Ntziachristos, “Looking at sound: optoacoustics with all-optical ultrasound detection,” Light: Science & Applications, vol. 7, no. 1, p. 53, 2018. View at: Publisher Site | Google Scholar
  19. E. Zhang, J. Laufer, and P. Beard, “Backward-mode multiwavelength photoacoustic scanner using a planar Fabry-Perot polymer film ultrasound sensor for high-resolution three-dimensional imaging of biological tissues,” Applied Optics, vol. 47, no. 4, pp. 561–577, 2008. View at: Publisher Site | Google Scholar
  20. J. A. Guggenheim, J. Li, T. J. Allen et al., “Ultrasensitive plano-concave optical microresonators for ultrasound sensing,” Nature Photonics, vol. 11, no. 11, pp. 714–719, 2017. View at: Publisher Site | Google Scholar
  21. G. Li, Z. Guo, and S.-L. Chen, “Miniature all-optical probe for large synthetic aperture photoacoustic-ultrasound imaging,” Optics Express, vol. 25, no. 21, pp. 25023–25035, 2017. View at: Publisher Site | Google Scholar
  22. B. Dong, C. Sun, and H. F. Zhang, “Optical detection of ultrasound in photoacoustic imaging,” IEEE Transactions on Biomedical Engineering, vol. 64, no. 1, pp. 4–15, 2017. View at: Publisher Site | Google Scholar
  23. A. P. Jathoul, J. Laufer, O. Ogunlade et al., “Deep _in vivo_ photoacoustic imaging of mammalian tissues using a tyrosinase-based genetic reporter,” Nature Photonics, vol. 9, no. 4, pp. 239–246, 2015. View at: Publisher Site | Google Scholar
  24. Y. Chen, B. Chen, T. Yu et al., “Photoacoustic mouse brain imaging using an optical Fabry-Pérot interferometric ultrasound sensor,” Frontiers in Neuroscience, vol. 15, 2021. View at: Publisher Site | Google Scholar
  25. Y. Liang, L. Li, Q. Li et al., “Photoacoustic computed tomography by using a multi-angle scanning fiber-laser ultrasound sensor,” Optics Express, vol. 28, no. 6, pp. 8744–8752, 2020. View at: Publisher Site | Google Scholar
  26. Y. Liang, L. Jin, L. Wang, X. Bai, L. Cheng, and B.-O. Guan, “Fiber-laser-based ultrasound sensor for photoacoustic imaging,” Scientific Reports, vol. 7, no. 1, 2017. View at: Publisher Site | Google Scholar
  27. R. Shnaiderman, G. Wissmeyer, O. Ülgen, Q. Mustafa, A. Chmyrov, and V. Ntziachristos, “A submicrometre silicon-on-insulator resonator for ultrasound detection,” Nature, vol. 585, no. 7825, pp. 372–378, 2020. View at: Publisher Site | Google Scholar
  28. R. Shnaiderman, Q. Mustafa, O. Ülgen et al., “Silicon-photonics point sensor for high-resolution optoacoustic imaging,” Advanced Optical Materials, vol. 9, no. 20, p. 2100256, 2021. View at: Publisher Site | Google Scholar
  29. W. J. Westerveld, M. Mahmud-Ul-Hasan, R. Shnaiderman et al., “Sensitive, small, broadband and scalable optomechanical ultrasound sensor in silicon photonics,” Nature Photonics, vol. 15, no. 5, pp. 341–345, 2021. View at: Publisher Site | Google Scholar
  30. H. Li, B. Dong, Z. Zhang, H. F. Zhang, and C. Sun, “A transparent broadband ultrasonic detector based on an optical micro-ring resonator for photoacoustic microscopy,” Scientific Reports, vol. 4, p. 4496, 2014. View at: Google Scholar
  31. A. Maxwell, S.-W. Huang, T. Ling, J.-S. Kim, S. Ashkenazi, and L. Jay Guo, “Polymer microring resonators for high-frequency ultrasound detection and imaging,” IEEE Journal of Selected Topics in Quantum Electronics, vol. 14, no. 1, pp. 191–197, 2008. View at: Publisher Site | Google Scholar
  32. T. Ling, S.-L. Chen, and L. J. Guo, “High-sensitivity and wide-directivity ultrasound detection using high Q polymer microring resonators,” Applied Physics Letters, vol. 98, no. 20, p. 204103, 2011. View at: Publisher Site | Google Scholar
  33. Z. Zhang, B. Dong, H. Li, F. Zhou, H. F. Zhang, and C. Sun, “Theoretical and experimental studies of distance dependent response of micro-ring resonator-based ultrasonic detectors for photoacoustic microscopy,” Journal of Applied Physics, vol. 116, no. 14, p. 144501, 2014. View at: Publisher Site | Google Scholar
  34. B. Dong, H. Li, Z. Zhang et al., “Isometric multimodal photoacoustic microscopy based on optically transparent micro-ring ultrasonic detection,” Optica, vol. 2, no. 2, pp. 169–176, 2015. View at: Publisher Site | Google Scholar
  35. H. Li, B. Dong, X. Zhang et al., “Disposable ultrasound-sensing chronic cranial window by soft nanoimprinting lithography,” Nature Communications, vol. 10, no. 1, p. 4277, 2019. View at: Publisher Site | Google Scholar
  36. M. Xu and L. V. Wang, “Universal back-projection algorithm for photoacoustic computed tomography,” Physical Review E, vol. 71, no. 1, article 016706, 2005. View at: Publisher Site | Google Scholar
  37. M. Mozaffarzadeh, A. Mahloojifar, M. Orooji, S. Adabi, and M. Nasiriavanaki, “Double-stage delay multiply and sum beamforming algorithm: application to linear-array photoacoustic imaging,” IEEE Transactions on Biomedical Engineering, vol. 65, no. 1, pp. 31–42, 2018. View at: Publisher Site | Google Scholar
  38. H. Kang, S.-W. Lee, E.-S. Lee, S.-H. Kim, and T. G. Lee, “Real-time GPU-accelerated processing and volumetric display for wide-field laser-scanning optical-resolution photoacoustic microscopy,” Biomedical Optics Express, vol. 6, no. 12, pp. 4650–4660, 2015. View at: Publisher Site | Google Scholar
  39. D. Lee, E.-Y. Park, S. Choi et al., “GPU-accelerated 3D volumetric X-ray-induced acoustic computed tomography,” Biomedical Optics Express, vol. 11, no. 2, pp. 752–761, 2020. View at: Publisher Site | Google Scholar
  40. A. Sun, H. Guo, Q. Gan, L. Yang, Q. Liu, and L. Xi, “Evaluation of visible NIR-I and NIR-II light penetration for photoacoustic imaging in rat organs,” Optics Express, vol. 28, no. 6, pp. 9002–9013, 2020. View at: Publisher Site | Google Scholar
  41. J. Xia, J. Yao, and L. V. Wang, “Photoacoustic tomography: principles and advances,” Electromagn Waves (Camb)., vol. 147, pp. 1–22, 2014. View at: Publisher Site | Google Scholar
  42. M. Xu and L. V. Wang, “Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction,” Physical Review E, vol. 67, no. 5, article 056605, 2003. View at: Publisher Site | Google Scholar
  43. M. Haltmeier and G. Zangerl, “Spatial resolution in photoacoustic tomography: effects of detector size and detector bandwidth,” Inverse Problems, vol. 26, no. 12, p. 125002, 2010. View at: Publisher Site | Google Scholar

Copyright © 2022 Qiangzhou Rong et al. Exclusive Licensee Suzhou Institute of Biomedical Engineering and Technology, CAS. Distributed under a Creative Commons Attribution License (CC BY 4.0).

 PDF Download Citation Citation
Views421
Downloads299
Altmetric Score
Citations