Research / 2022 / Article

Research Article | Open Access

Volume 2022 |Article ID 9870637 | https://doi.org/10.34133/2022/9870637

Yiqun Liu, Qi Yu, Xiaojin Luo, Le Ye, Li Yang, Yue Cui, "A Microtube-Based Wearable Closed-Loop Minisystem for Diabetes Management", Research, vol. 2022, Article ID 9870637, 14 pages, 2022. https://doi.org/10.34133/2022/9870637

A Microtube-Based Wearable Closed-Loop Minisystem for Diabetes Management

Received29 May 2022
Accepted12 Sep 2022
Published27 Oct 2022

Abstract

Diabetes is a chronic metabolic disease with a high blood glucose level, leading to both seriously acute and chronic complications. The closed-loop system is an ideal system for diabetes management. However, the large size and high cost of the commercial systems restrict their widespread uses. Here, we present for the first time a microtube-based wearable closed-loop minisystem for diabetes management. The closed-loop minisystem includes a biosensing device, an electroosmotic micropump, and a printed circuit board (PCB) with an algorithm. The microtube-based sensing device coated on the outer surface of the microtube is inserted into subcutaneous tissue for detecting interstitial glucose; the current signal for sensing glucose is processed by the PCB to power the electroosmotic micropump intelligently for the delivery of insulin into the subcutaneous tissue via the microtube channel. The closed-loop minisystem worn on a diabetic SD rat can successfully maintain its blood glucose level within a safe level. It is expected that this new closed-loop paradigm could open up new prospects for clinical diabetes management.

1. Introduction

Diabetes, as a global chronic disease among millions of people [1, 2], is caused by the inability to produce insulin because of pancreatic dysfunction (type I diabetes) [3] or caused by the malfunction of cells to use insulin (type II diabetes) [4]. Diabetes management is highly important for diabetes patients, since a lack of proper management could result in serious complications, such as heart disease, stroke, kidney failure, blindness, and nerve damage [57]. The key for a desired management is to control the blood glucose at a normal level, usually in the range of less than 7.0 mM [8].

Diabetes patients usually need three to four blood glucose concentration samples per day obtained by the patient by using fingerstick glucose meters, which is painful and inconvenient [9]. When blood glucose is higher than a normal level, patients would take oral drugs (such as insulin secretagogues [10], noninsulin secretagogues [11], DPP-4 inhibitors [12], and SLGT-2 inhibitors [13]) or subcutaneously inject insulin. These ways are inconvenient for the patients and always result in a poor control of diabetes. To improve the user experience and the working performance, novel wearable biosensors and drug delivery devices have been studied, such as the flexible self-sustainable sweat sensors [14], the integrated watch for continuous glucose monitoring [15], and the multimicrochannel microneedle for intracellular drug delivery [16]. Recently, the closed-loop devices have attracted great interest due to their intelligent function for automatic monitoring blood glucose and injecting insulin, such as the system developed by Medtronic Inc. (Northridge, CA) [17]. The device normally includes a continuous glucose meter and a pump in separated parts, and the pumping amount of insulin is adjusted by the glucose level, which is controlled by an electronic chip with algorithm. Therefore, the closed loop is usually very large in size, and it normally has a dimension of  cm.

The intradermal route is feasible for sensing glucose and delivering insulin [1821]. Several studies have been performed to develop the closed-loop devices towards the dermis layer, such as a nanosensor array on microneedle elements for monitoring glucose levels and delivering insulin [22], the wearable patch based on a sweat biosensor and polymeric thermoresponsive microneedles [23], and a system based on mesoporous microneedles and iontophoresis technology [24]. We have recently shown a microneedle sensing device with an electroosmotic micropump for a closed-loop diabetes management [25] as well. However, the microneedles can only reach the dermis layer, and these devices suffer from some key disadvantages, such as the unreliability of sensing components, the imprecise control of the insulin release rate, and the leakage of insulin due to the small height of the microneedles. In addition, the capillaries in the dermis are not abundant which may limit the exchange rate of the glucose molecules between cells and blood vessels to result in a time delay for glucose sensing.

Subcutaneous tissue has more capillaries to result in an enhanced exchange rate of glucose molecules between cells and blood vessels. The time lag between glucose fluctuation in subcutaneous tissue and glucose fluctuation in blood is shorter than that in the dermis layer. In addition, insulin can be absorbed faster in the subcutaneous tissue than that in the dermis. There is no insulin accumulation in the subcutaneous tissue or insulin leakage out of the skin.

A microtube is usually made of a flexible polymer [26, 27], or a metal [28]. Compared with a metal, a polymeric microtube has a high biocompatibility, causes less pain, and can control the implanting angle conveniently [29, 30]. A microtube with a length of several millimeters could be implanted into subcutaneous tissue, usually for delivering a drug, such as insulin [31], tenofovir alafenamide [32, 33], dexamethasone [34], and estradiol-17 beta [35].

In this work, we show for the first time a flexible, biocompatible microtube-based closed-loop minisystem for diabetes management. The outer surface at one end of the microtube is coated with a biosensing device to detect the glucose levels in subcutaneous interstitial fluid continuously. The other end of the microtube is integrated with an electroosmotic micropump that can provide a precise insulin delivery into subcutaneous tissue through the microchannel of the microtube. A printed circuit board (PCB) is connected to both the sensor and the micropump, to power the sensor, detect the sensing signal, and drive the micropump to deliver the insulin according to the sensing signal from subcutaneous glucose. The microtube enables an effective injection of insulin into subcutaneous tissue without a leakage or a low adsorption. The biosensing device on the outer device with a length along the entire dermis and subcutaneous tissue enables an effective sensing of glucose. The closed-loop device is wearable, small, and cost-effective.

2. Results and Discussion

2.1. Overall Device Principle and Components of the Closed-Loop System

As illustrated in Figure 1(a), the system included a glucose biosensor with two electrodes deposited on the opposite sidewalls of the microtube, an electroosmotic insulin micropump, and a flexible catheter to connect them. The flexible microtube with a diameter of 400 μm and a length of 6 mm was inserted into the subcutaneous tissue transdermally, passing through the epidermis (75–150 μm) and dermis layer (1–4 mm) [36]. The Au and Ag/AgCl electrode was able to provide an electrochemical sensing of glucose. This transdermal sensing approach for glucose has been proven to be feasible, since many glucose molecules were transferred from capillaries to interstitial fluid when the blood glucose level changes rapidly. The nanoporous polycarbonate membrane was the core of the micropump to achieve an electroosmotic flow. Two stainless-steel meshes with an Au-layer deposition were adhered to both sides of the membrane as the anode and cathode of the micropump. Before placing the system on a skin, the skin surface at the penetration position was disinfected by ethanol without needing a local anesthesia. The system was then worn on the skin (such as abdomen and upper arm) and penetrate the skin surface into the subcutaneous tissue.

Figure 1(b) illustrates the working principle of the system. The microtube was inserted into the arm of the human body to perform a glucose sensing and an insulin release. Glucose was catalyzed by glucose oxidase (GOD) to produce H2O2 that was further detected by the sensing electrode to result in a current response. A high blood glucose level triggered the PCB automatically to apply a constant potential on the electroosmotic micropump for releasing insulin through the hollow microtube. With the injection of insulin into subcutaneous tissue, the blood glucose level gradually declined to a normal range. The system could be further connected with a wireless unit for transmitting the blood glucose data to a portable device (e.g., smartphone), and users could know the fluctuations of blood glucose conveniently.

Figure 1(c) shows the camera images of the microtube biosensor. The microtube was flexible and can penetrate into the skin with the assistance of a stainless-steel puncture guide needle core. Figure 1(d) shows the SEM images of the flexible microtube. The tip of the microtube was in a trapezoid shape, and the center of the microtube was in a cylinder shape. The microtube was 6 mm in length and 0.4 mm in diameter with a sidewall thickness of about 50 μm. Figure 1(e) exhibits a photograph of a polycarbonate electroosmotic micropump with two Au-deposited stainless-steel meshes adhered on its two sides as the electrodes. The diameter of the circular polycarbonate membrane was 2.5 cm with a thickness ranging from 7 to 22 μm. The rectangular mesh was 1 cm in width and 2.5 cm in length with a thickness of about 100 μm. A 3D-printed hemispherical-shaped drug reservoir (2 cm in diameter) was adhered with the micropump for storing the insulin solution. There were multiple pores (200 nm in diameter) on the surface of the polycarbonate membrane to achieve an electroosmotic flow (Figure 1(f)).

The entire system can be very small and wearable with a centimeter size (Figure S1). The microtube can be inserted into the rat skin successfully (Figure 1(g)). After removing the system, there was no bleeding and severe traumatic wound in the skin. Therefore, there was no serious injury to the skin in using the system.

A printed circuit board (PCB) was used to control the sensing, pumping, and feedback processes for the biosensor and micropump (Figure S2). The PCB can provide the constant potentials for the biosensor (0.1–0.3 V) and the micropump (0–10 V), receive data from the biosensor, transmit the data to the computer, and control the micropump for the insulin release. From the schematic diagram and circuit diagram of the PCB for the biosensor and the micropump (Figures S2 and S3), the current measured by the biosensor was processed by the PCB that determined whether the blood glucose level was higher than the critical value. If the blood glucose was higher, it would drive the micropump to release the insulin solution. After a short period of the insulin release, the PCB would drive the biosensor to perform a glucose sensing again. The sensing and insulin delivering processes would be repeated alternatively until the blood glucose level reached the normal level. The design of this PCB was similar to the one we reported previously [25].

2.2. Construction of Biosensor for Detecting H2O2

Figure 2(a) illustrates the fabrication process of the biosensor. Two opposite sidewalls of the microtube were deposited with Ti/Au as the working electrode and a base layer for the Ag electrode (1-2). A solid-state Ag/AgCl was designed for being the counter/reference electrode, and it was formed by an Ag layer deposition and its chlorination (3-4). A thin-film Prussian blue (PB) layer was then electrodeposited on the Au working electrode (5), as a mediator for H2O2 to transfer the electrons to the working electrode. It can ensure the biosensor being operated at a low potential (0.1 V), extend the detection range, and increase the sensitivity. GOD was then immobilized on the Au electrode to catalyze the glucose molecules to generate H2O2 for sensing (6). The biocompatible chitosan/Nafion membrane was then deposited as the encapsulation matrix for GOD to protect the enzyme from leakage, control glucose diffusion, eliminate interferences, and enhance the biocompatibility of the device (7-8).

As shown in Figures 2(b) and 2(c), each electrode had a width of 0.4 mm and a height of 6 mm, and two electrodes were successfully fabricated on the sidewall of the microtube. The Au electrode was in a yellow color (Figure 2(b)), the Ag/AgCl electrode was formed by chloridizing the Ag layer and in a gray color (Figure 2(c)). After the deposition of PB, the working electrode showed a slight blue color (Figure 2(d)). A stainless-steel needle that was longer than the microtube would guide the microtube to be inserted into the skin. Although the insertion with a stainless-steel guide needle may cause a little pain to the users, the microtube was soft, flexible, and very thin (only 400 μm in diameter) that would minimize the pain feeling after the insertion. It has been demonstrated that the pain was almost the same when using a microneedle and a subcutaneous microtube for the insulin infusion [37]. A variety of microtubes with different dimensions and shapes can be used, and a 400 μm-diameter microtube can also be replaced by a microtube with a smaller diameter [38].

The characterizations by SEM and EDS was performed on a bare Au electrode, an Ag/AgCl electrode, a PB-deposited Au electrode, and a GOD and Nafion-deposited electrode, as can be seen in Figures 2(e) to 2(g) (Figure S4 to Figure S5). The results indicate that Au and Ti elements were distributed on the sidewall of the microtube evenly. Figure 2(f) and Figure S5b also indicate that Ag and Cl elements were also evenly distributed on the Ag/AgCl electrode. Compared to a bare Au electrode, the N and Fe elements were distributed well on the surface of the Au electrode, demonstrating the successful deposition of PB (Fe4[Fe(CN)6]3) on the Au electrode (Figure 2(g) and Figure S5c). The Nafion ((C7HF13O5S·C2F4)x) membrane was the outmost layer of the Au electrode; the EDS mapping indicates that the S and F elements were also distributed on the Au electrode evenly (Figure S4 and Figure S5d). There were some cracks in the Nafion membrane that may enhance the diffusion of O2 and glucose molecules into the enzyme layer.

Figure 2(h) shows the cyclic voltammograms (CV) of the PB-deposited Au electrode in a 0.1 M KCl/HCl solution. A CV scanning was performed on the sensor to deposit PB with 4 cycles each time for twice. The CV performances were stable in the two repeated scanning experiments for depositing PB. After the PB modification, the sensor was immersed into a KCl/HCl solution for a CV scanning from −0.2 V to 0.5 V to stabilize the PB layer. Figure 2(i) shows the current responses of the biosensors with different thicknesses of the PB layers to 4 mM H2O2. The sensing response with a PB deposition for 8 cycles was faster and larger that was chosen for the further experiments. After the PB deposition, the electron transfer impedance of the electrode declined sharply (Figure S6). Figure 2(j) shows the CV curves of the sensor with 4 mM H2O2 in PBS. The peak currents of the CV curves were proportional to the square root of the scanning rates, indicating a diffusion-controlled process for the H2O2 sensing [3941]. The current values were variable at different potentials due to the different reduction and oxidation rates of H2O2. A 0.1 V potential was chosen for the further amperometric experiments due to the clear signal response in the CV curves. Figure S7 displays the current response of the sensor for detecting H2O2 in PBS and the calibration curve between the sensing signal and the H2O2 concentrations. A linear detection range was clearly obtained from 0.8 mM to 85 mM with a sensitivity of μA/mM (). The detection limit was calculated to be 13.78 μM (signal to noise was 3). H2O2 is the product of glucose molecules oxidized by the GOD, and the successful detection of H2O2 indicates the possibility to measure glucose accurately.

2.3. Electrochemical Sensing of Glucose with the Microtube Sensor

Figure 3(a) shows the CV curves of the sensor to 4 mM glucose in PBS with a potential range from −0.5 V to 0.5 V. By changing the potentials, the current values varied, and a linear relationship can be seen between the peak current and scanning rate, indicating a diffusion-controlled process for sensing glucose as well. Figure 3(b) shows the sensing responses to a series of glucose additions in PBS. The biosensor was able to detect a wide glucose concentration range (0.8–34 mM) with a sensitivity of μA/mM (). These concentrations could cover the entire blood glucose levels in the diabetic patients as well as the normal persons. Figure 3(c) shows the current baseline response of the sensing device to glucose in PBS. Its performance was similar to that in Figure 3(b) with a sensitivity of μA/mM and a linear detection range of 0.8–34 mM (). Figure 3(d) illustrates the currents to different glucose concentrations in simulated interstitial fluid and the calibration curve for the sensor. The calibration curve exhibited a slope of μA/mM and a 0.8–34 mM linear range ().

Figure 3(e) shows the selectivity of the sensing device. The selectivity of the biosensor is essential to achieving a high accuracy in detecting interstitial glucose, since many electroactive substances could exert the interfering effects on the glucose sensing. The results showed that the biosensor was more selective to glucose than other electroactive interferences, including uric acid, ascorbic acid, dopamine, and insulin. Figure 3(f) shows the current-versus-time curve of the microtube sensing device after adding different volumes of insulin (10 U/μl) and glucose (4 mM) in PBS. Due to a continuous injection of insulin, the delivered insulin may accumulate around the sensor and affect its accuracy. These results demonstrate that different volumes of insulin almost had no effect on the sensing response to glucose.

In addition, the sensing performance was almost unaffected upon applying different bending angles and bending times on the microtube due to its great flexibility, as can be seen Figures 3(g) and 3(h). The relative response after the microtube being bent for 90°once was 91.69%, and it was 88.67% with a 200 times’ bending at 45°. The operational, pH, temperature, and storage stabilities were evaluated, as shown in Figures 3(i) to 3(l). The sensor exhibited the excellent performances (above 80% of the initial response) for over fifty continuous measurements, for sensing at different temperatures, for sensing at different pH values, and for over 7 days. The relative responses of the biosensors at pH 6.5, at 20°C, and at the 7th day were 88.31%, 84.43%, and 84.78%, respectively. All these results demonstrate that the biosensor is sensitive and stable for detecting glucose under different circumstances.

2.4. Characterization of the Electroosmotic Micropump

Figure 4(a) shows the working principle of an electroosmotic micropump for insulin. The nanopore polycarbonate membrane was the key component of the micropump. Two stainless-steel meshes after the Au deposition was attached with the membrane as electrodes, and a constant DC voltage was applied to electrodes. During its working process, the surface charges from the ionizable groups or a strong adsorption of the charged species were accumulated at walls of the pores. Therefore, the oppositely charged ions from the electrolyte were attracted by these charges to form an electric double layer. Due to a thermal diffusion, a diffusion layer with a concentration gradient of the oppositely charged ions was formed near the walls. By applying an electrical field in the axial direction, the electric force would drive the net charges in the diffusion layer towards the oppositely charged electrode, dragging the solvent along to produce an electroosmotic flow. Figure 4(b) shows the camera image of a micropump. A 3D-printed button shape chamber was adhered to the top of the membrane as the insulin reservoir, and the micropump was connected to a microtube via a medical-grade PTFE hollow tube (4 mm in diameter and ~2 cm in length) in between. The micropump was connected to a PCB, which provided a constant voltage with a range of 0 to 10 V to power the micropump. The PCB was further connected to a computer through a USB cable to transmit the current and voltage data.

Figure 4(c) shows the rate of the electroosmotic micropump at a DC voltage of 10 V for pumping different concentrations of insulin. With an increasing insulin concentration from 1 U/ml to 20 U/ml, the flow rate gradually decreased from 7.5018 μl/min to 1.5198 μl/min, probably due to the elevated viscosity from a high concentration of insulin. Figure 4(d) shows the change of flow rates at different DC voltages for pumping 10 U/ml and 20 U/ml insulin solutions. The flow rate was linearly increased from 1.8393 μl/min (10 U/ml) and 0.6189 μl/min (20 U/ml) to 3.376 μl/min (10 U/ml) and 1.5198 μl/min (20 U/ml) when the potential was increased from 1 V to 10 V. Users could further change the rate of insulin release by adjusting the DC voltage applied to the micropump. Figure 4(e) shows the current-versus-time curve at different potentials for releasing insulin solution (10 U/ml). A larger current was obtained at a higher potential, and the current at each voltage was relatively stable during the measuring time (60 s). Figure 4(f) exhibits the power and power/flow rate for micropump at different potentials. The power needed by the micropump was very low (3.16 mW at 10 V) for delivering insulin (10 U/ml) at a flow rate of 3.376 μl/min. In addition, the ratio of power/flow rate increased with an increasing potential, and this indicates that the micropump was more efficient at a lower potential. All these results indicate that the micropump exhibited an excellent performance for delivering insulin.

2.5. In Vivo Performance of the Closed-Loop System

Before conducting the closed-loop study, normal Sprague Dawley (SD) rats were induced to be diabetic rats by injecting STZ according to the protocol [42]. The closed-loop system was applied on the back skin of a diabetic rat. The microtube sensing device and the electroosmotic micropump were connected to a PCB that exchanged data with a PC (Figure 5(a)). Figure 5(b) shows a camera image of the system applied to an SD rat. The back skin of the rat was selected as the insertion site of the system, and after removing hairs, the microtube was plunged into the skin transdermally. The microtube device was successfully inserted into the rat’s skin that was demonstrated from a hematoxylin and eosin-stained pierced skin section (Figure S8a). After applying a 20-gram object on the skin, the rat’s skin was almost not changed (Figure S8b). The results indicate that due to its flexibility and softness, the microtube device would not cause serious tissue damage in case of a fall of a wearer. From the camera images of the rat’s skin after applying a microtube device for 5 days (Figure S9), there was no obvious skin reaction, such as serious erythema and edema, demonstrating an excellent biocompatibility and safety of the microtube device. After removing the microtube device, the more pores left on the skin disappeared in about four days (Figure S10).

A two-step mode was adopted to achieve a closed-loop management, as illustrated in Figure 5(c). The red color is the current-versus-time () curve for sensing glucose (the left line corresponded to 13.1 mM blood glucose, and the right line corresponds to the 8.2 mM). After measuring glucose for 50 s, the final current value of the curve was recorded and sent to the PCB to decide if the blood glucose level reached the critical value (8.3 mM). If the value was higher than the critical value (8.3 mM), a 5 V potential was applied to the micropump to inject insulin (10 U/ml) for 600 s. The sensing and pumping steps were repeated alternatively until the blood glucose was back to the normal value. After that, the insulin release was stopped and only the glucose sensing process was performed. When the pump was stopped, a 0.1 V potential was provided to the pump to prevent the loss of its flow rate caused by the repeated on/off micropump.

For the automatic closed-loop management of diabetes, the first step is to obtain the relationship between the current measured by the biosensor and the true blood glucose level. As shown in Figure 5(d), the current change from the microtube biosensor correlated well with the blood glucose level measured by a commercial glucometer, and a linear relationship was seen with a sensitivity of 0.0901 μA/mM (Figure 5(e)). All 135 points from six different rats were located in the safe error zone (A and B) of the Clark error grid (Figure 5(f)). The error of the biosensor was from 0.094% to 25.33%, The mean absolute relative difference (MARD) value of the |relative error| between these two detection approaches was , and 75% of points were lower than 14.241% (Figure S11). The results fulfilled the criteria for the accuracy in glucose sensing [43]. All these results indicate that the biosensor could measure the blood glucose level accurately.

Figure 5(g) shows the changes in the blood glucose levels with no treatment, a microtube injection, a saline delivery, an insulin delivery, and the one-time subcutaneous injection of insulin. No decrease in the blood glucose levels can be seen with no treatment, a microtube injection, or a saline delivery. When the microtube was applied to a rat and insulin was released by a pump, the blood glucose level was lowered to 45% of its initial value in two hours. In addition, the hypoglycemic ability of the system was similar to the one-time subcutaneous injection of the same amount of insulin in 130 min (the amount of insulin was calculated according to the flow rate of the micropump).

Figure 5(h) shows the closed-loop management of diabetes without a glucose intake. The blood glucose of the first rat (blue line) was declined dramatically in about two hours. At the time of 130 min, when the level of blood glucose decreased from 19.3 mM to 8.2 mM (lower than the critical value of 8.3 mM), the insulin release was stopped automatically. Without a further insulin intake, the blood glucose level increased again and reached 11.4 mM at 240 min. For the second rat (red line), the blood glucose decreased from 20.8 mM to 8.2 mM in two hours. The insulin release was also stopped automatically, and the blood glucose increased to 8.8 mM at the time of 140 min. Then, the pump was driven again automatically. Gradually, with the increase in the amount of insulin, the blood glucose reached a peak of 9.9 mM at the time of 180 min, and after that, the blood glucose began to decrease until it reached 8.2 mM at the time of 210 min.

Figure 5(i) shows the closed-loop management of diabetes with a glucose intake. Similarly, the blood glucose level of the first rat (blue line) declined from 20.5 mM to 8.2 mM (lower than the critical value of 8.3 mM), and at the time of 130 min, the insulin release was then stopped automatically. At the time of 150 min, glucose (1 g/kg) was injected intraperitoneally. Without any operation, the blood glucose level increased gradually and reached 14.0 mM at the time of 240 min. For the second rat (red line), the blood glucose level decreased from 21.1 mM to 8.2 mM at 130 min. The insulin release was also stopped automatically. At the time of 140 min, the glucose (1 g/kg) was also injected intraperitoneally. The blood glucose level increased to 8.4 mM at the time of 150 min (higher than 8.3 mM). Then, the pump was turned on again automatically. Gradually, with the increase in the amount of insulin, the blood glucose level reached a peak of 10.5 mM at the time of 180 min. After that, the blood glucose level began to decrease until it reached the critical value (8.3 mM) at the time of 240 min. All these results demonstrate that the system could achieve an automatic closed-loop control of blood glucose in diabetic rats successfully.

Compared to other work for the closed-loop diabetes systems based on microneedles [2225, 44] (Table S1), this work exhibit several obvious advantages. It can deliver insulin into subcutaneous tissue with a precise volume control that can differentiate a volume as low as 0.03 μl. Insulin in subcutaneous tissue can be adsorbed faster without an insulin leakage. In addition, although there is only a microtube with a diameter in micrometers being inserted into the skin instead of arrays with a millimeter to centimeter size, the microtube sensing electrodes along the dermis layer and subcutaneous tissue can provide a sufficient electrode area to contact interstitial fluid for a sensitive glucose sensing. The device has a low cost and a small dimension as well.

In practical applications, a strong bandage or medical tape can be used to fix it on the skin to reduce the effect from a motion or an external force. The accuracy of the sensor is affected by a change in pH and temperature, and to improve the accuracy, the system can be further integrated with a pH or temperature sensor to calibrate the sensing results. Further, the insertion site can be covered by a waterproof membrane to prevent sweat from contacting the microtube to affect the accuracy. The system is disposable due to its low cost, and for a long-term diabetes management, the users can replace the microtube every few days to avoid the block of the microtube and the insulin infusion by fat granules, and other possible complications [45].

This work mainly focused on the closed-loop device parts (more specially, the sensing and pumping parts), since these are critical for the miniaturization and wearability of the closed-loop device. Although the developments of the PCB and the control algorithm are important as well, these are not the focus of this study. Therefore, this study used a simple designed PCB and control algorithm for operating the closed-loop device, which certainly should be improved in the future by the professional PCB and microprocessor engineers. Although this new closed-loop device is promising for clinical use, more specialized electronic chips, PCBs, and control algorithms should be designed and used eventually for the practical applications on human diabetic patients.

3. Conclusions

Here, we have successfully demonstrated an integrated closed-loop diabetes management system that consists of a microtube-type biosensor implanted into the subcutaneous tissue to detect glucose, an electroosmotic micropump to inject insulin, and a PCB to achieve the control algorithm. The biosensor was constructed with a two thin-film electrode configuration on the microtube, and the electroosmotic micropump was based on a low-cost nanoporous polycarbonate membrane. In vivo experiments proved that the system could perform a closed-loop control of blood glucose in diabetic rats intelligently. Compared to other commercial closed-loop systems, the system is small, lightweight, and cost-effective. Although these results are promising, further studies are needed to redesign the PCB to equip it with a wireless function, enhance the control algorithm, and evaluate the capability of the closed-loop system for managing blood glucose in human diabetic patients. This work can open a new paradigm for developing new closed-loop diabetes devices and may provide exciting opportunities for widespread uses among diabetes patients.

4. Materials and Methods

4.1. Apparatus and Chemicals

A potentiostat CHI660e was acquired from CH Instruments, Inc. (Shanghai, China). The magnetron sputtering was obtained from Kurt J. Lesker Co., Ltd. (America). The UV Ozone Cleaner was obtained from Ossila Co., Ltd. (Sheffield, UK). The hot plate was obtained from Jingxue Scientific Instrument Co., Ltd. (Shanghai, China). The S-4800 field emission scanning electron microscope was obtained from HITACHI, Ltd. (Tokyo, Japan). The SLA550 3D printing machine was obtained from ZRapid Tech Co., Ltd. (Suzhou, China). The commercial blood glucose meter was obtained from Sinocare Inc. (Changsha, China).

The flexible microtube was obtained from Juxin Anchuang Medical Device Co., Ltd. (Shenyang, China). The polycarbonate membrane with a diameter of 2.5 cm for constructing the electroosmotic micropump was purchased from Whatman Co., Ltd. (Meterstone, UK). The 304 stainless-steel mesh was purchased from the Sheng Cong Screen Operation Department (Changzhou, China).

Sodium hydroxide was obtained from Beijing Chemical Works (Beijing, China). Sodium dihydrogen phosphate was obtained from Tianjin Fuchen Chemical Reagent Factory (Tianjin, China). Glucose oxidase (GOD) was obtained from Toyobo Ltd. (Osaka, Japan). Glucose, glutaraldehyde, dopamine hydrochloride, and streptozotocin (STZ) were obtained from Sigma-Aldrich Inc. (Beijing, China). Ferric trichloride, calcium chloride anhydrous, potassium chloride, and hydrochloric acid were obtained from Xilong Science Co., Ltd. (Shantou, China). Potassium ferricyanide was obtained from Aladdin Inc. (Beijing, China). Sodium alginate and chitosan were obtained from Sinopharm Chemical Reagent Co., Ltd. (Shanghai, China). Sulfuric acid was obtained from Beijing Institute of Chemical Reagents (Beijing, China). Nafion solution (5%) was purchased from D&B Biological Science and Technology Co. Ltd. Ascorbic acid and uric acid were purchased from Meryer Chemical Technology Co., Ltd. (Shanghai, China). The fast-acting insulin aspart and long-acting detemir injection solution were obtained from Novo Nordisk Pharmaceutical Co., Ltd. (Beijing, China).

4.2. Sensing Electrode Preparation

The biosensor was fabricated with an Au working electrode and an Ag/AgCl electrode The flexible microtube was first cleaned in the UV ozone for 10 minutes. To fabricate the Au electrode, a 20 nm-thick Ti layer and a 200 nm-thick Au layer were deposited on the microtube to form two electrodes. For the Ag/AgCl electrode, a 200 nm-thick Ag layer was deposited on one Au electrode and then immersed into 50 mM ferric chloride (FeCl3) for 10 s to form the AgCl layer. The distance between the two electrodes on both sides was about 1 mm. Then, a CV was performed on the sensing electrode in 0.1 M H2SO4 at a rate of 1 V/s from 0.2 V to 1.2 V for 10 cycles to remove the impurities from the electrode surface and activate the electrodes. Then, the microtube was immersed into a solution containing 2.5 mM FeCl3, 100 mM KCl, 2.5 mM K3Fe(CN)6, and 100 mM HCl for electrodepositing the PB mediator layer onto the Au working electrode. The CV scanning was conducted at a rate of 20 mV/s from −0.15 V to 0.3 V (vs. Ag/AgCl) for 8 cycles. Finally, the biosensor was immersed into a 0.1 M KCl/HCl solution to repeat the CV scanning at 50 mV/s from −0.2 V to 0.5 V (vs. Ag/AgCl) for 4 cycles to further stabilize the PB layer onto the Au electrode.

4.3. Enzyme Functionalization

The sensing electrodes were firstly treated using a UV Ozone Cleaner for 10 min to obtain a hydrophilic surface before immobilizing the enzymes to bring a good contact of the enzyme with the Au electrode surface. Then, the GOD solution (50 U/μl) was mixed with 1% BSA solution at a volume radio of 1 : 1, and 10 μl of the mixture was coated on the surface of the Au electrode. 10 μl of diluted glutaraldehyde (2%) was coated on the electrode. After drying at the 4°C for 2 h, 5 μl of the chitosan solution (1%, dissolved in the 2% acetic acid) was coated onto the Au electrode. After waiting for 2 h to dry the chitosan layer, the biosensor was modified with a 5 μl Nafion (0.5%) solution. The sensor was then stored in a refrigerator at 4°C overnight.

4.4. Calibration of H2O2 and Glucose Measurement

A constant potential of 0.1 V (vs. Ag/AgCl electrode) was applied to the working electrode of the sensor by a potentiostat to obtain the sensing current signal at room temperature (about 24°C). The sensing of glucose was studied in PBS (50 mM pH 7.0) and simulated interstitial fluid. In PBS, the biosensor was immersed into 150 μl PBS, and a series of 6 μl of H2O2 or glucose with different concentrations were added into the solution incrementally.

To prepare the simulated interstitial fluid, alginic acid sodium salt was first slowly added to 0.1 M KCl to produce a 1.5% w/v viscous solution. The mixture was stirred at 45°C for 4 hours in the oven. After that, a series of solid glucose with different qualities were dissolved into the mixture to generate the required concentration solution. Then, 0.2 M calcium chloride (CaCl2) was added until the alginate solution was completely covered. Finally, the solution was covered with parafilm and left in a refrigerator at 4°C to induce complete crosslinking. During the detection process, the biosensor was stuck into the hydrogel with different glucose concentrations.

4.5. Selectivity Test of the Biosensor to Different Interferences

To study the interfering effects of different electroactive substances on the sensor, a series of 6 μl of uric acid, ascorbic acid, dopamine, insulin, and glucose were added into PBS with a final concentration of 0.1 mM, and different volumes (100–400 μl) of insulin (10 U/μl) were added into PBS as well. The current-versus-time curve was recorded during the study.

4.6. pH, Temperature Influence, and Storage Stability Test

The effect of different pH values was tested with PBS at different pH values. The sensing response to 4 mM glucose was studied in buffer at pH 6.0, 6.5, 7.0, 7.5, or 8.0. The effect from the temperature was studied with a hot plate. The sensing response to 4 mM glucose was evaluated at different temperatures. To measure the storage stability of the biosensor, the sensing response to 4 mM glucose was studied for three times each day. The biosensor was immersed in PBS and stored in the refrigerator at 4°C during the free time.

4.7. Bending Time and Angle Influence

To evaluate different bending times on the sensing performance, the microtube was bent for 50, 100, 150, and 200 times at an angle of 45°, and the current responses to 4 mM glucose were measured. The sensing responses to 4 mM glucose were also measured when the microtube was bent at the angles of 30°, 45°, 60°, 75°, and 90° to study the effect of different bending angles.

4.8. Electroosmotic Micropump Preparation and Flow Rate Test

The electroosmotic micropump was composed of a polycarbonate membrane with a diameter of 2.5 cm and two 304 stainless-steel meshes. These meshes with a width of 1 cm and length of 2 cm were then deposited with a Ti (20 nm)/Au (200 nm) layer to construct the Au electrode. The Au electrode could increase the mesh conductivity of and prevent it from corrosion. The two meshes covered the polycarbonate membrane completely, and part of the mesh was extended outside for connecting to the external powering device. Finally, the catheter connecting to the flexible microtube was attached to the bottom of the membrane. A 3D-printed button shape chamber with a diameter of 2 cm, a height of 1 cm, and a wall thickness of 3 mm was adhered to the top of the micropump for storing insulin. The flow rate test was studied with an electronic balance (resolution 0.01 mg). Different amounts of liquids being delivered from the micropump were collected every 5 min, and the weight was measured by the balance. Then, the wight was converted to the volume according to the density of each insulin solution to calculate the flow rate.

4.9. Animal Experiments

All SD rats for experiments were purchased from Charles River Biotechnology Co., Ltd. (Beijing, China). The SD rats were male with a weight of 150–200 g and 8 to 10 weeks of age. All experimental procedures were approved by the Research Ethics Committee of Peking University First Hospital (Approval Number: 202061). Streptozotocin (STZ) was used to induce the rats to be diabetic. All rats were fasted for 6 to 8 h before STZ treatment, and only water was normally provided. Solid STZ was dissolved in the 50 mM sodium citrate buffer (pH 4.5), and the STZ solution was intraperitoneally injected into rats in a single dose of 65 mg/kg (according to the weight of rats). During the first day after STZ treatment, 10% sucrose water was provided to each rat, and since the second day, regular water was provided to each rat. All rats were stored in specific pathogen-free conditions, and they were maintained in individually ventilated cages during the intermission period of the experiment. The condition had a constant temperature of °C, a constant humidity of , and a 12 h light/dark cycle. The rats had access to standard laboratory food and water during free time. On the tenth day, all rats were fasted for 6–8 h, and the blood glucose level from a tail vein sample of each rat was tested to check hyperglycemia. If the value was higher than 8.3 mM, the rat was considered to be diabetic.

Before the experiments, to test the performance of the closed-loop system, all diabetic rats were fasted for 8 h and were then anesthetized with the assistance of a small animal anesthesia machine during experiments. The hair of the rat’s back in the insertion site was shaved and cleaned. As the next step, the system was stuck to the rat’s skin and waited for ~30 minutes for the biosensor to warm up and get stabilized in the subcutaneous before experiments. The insulin aspart Injection solution (10 U/ml) powered by an electroosmotic micropump under a 5 V DC voltage was subcutaneously injected into the rat through the microtube. To measure the change in the rat’s blood glucose level at different times, the blood sample was collected from the rat’s tail vein and was measured by the commercial glucose meter.

Data Availability

The data used to support the findings of this study are included within the article and the supplementary information file.

Conflicts of Interest

The authors declare that there is no conflict of interest regarding the publication of this article.

Authors’ Contributions

Y. Cui and L. Yang conceived and designed the study. Y. Liu, Q. Yu, and X. Luo led the experiments. L. Ye designed the PCB. All authors contributed the data analysis and interpretation. Y. Liu, Q. Yu, L. Yang, and Y. Cui wrote the paper, and all authors provided feedback. Y. Liu and Q. Yu contributed equally to this work.

Acknowledgments

The authors greatly acknowledge the National High Level Hospital Clinical Research Funding (High Quality Clinical Research Project of Peking University First Hospital, No. 2022CR83), National Natural Science Foundation of China (No. 52072007 and No. 82130021), National Key Research and Development Program of China (No. 2018YFB1307301), CAMS Innovation Fund for Medical Sciences (2019-I2M-5-046 and 2020-JKCS-009), PKU-Baidu Fund (2020BD026 and 2020BD044), Capital’s Funds for Health Improvement and Research (CFH2022-1-4071), and Beijing Young Scientist Program (BJJWZYJH01201910001006) for the financial support.

Supplementary Materials

Figure S1. (a) A camera image of the whole system. (b) A camera image of the system applied to the human body. Figure S2. Illustration of the working principle and control algorithm of the closed-loop system. Figure S3. Circuit diagrams of the PCB for the biosensor and micropump Figure S4. SEM image and EDS mapping analysis of the Au electrode after the immobilization of GOD and the Nafion deposition Figure S5. EDS point analysis of the Au electrode and the Ag/AgCl electrode. (a) The original Au electrode; (b) The Ag/AgCl electrode. (c) The Au electrode after the deposition of Prussian blue; (d) The Au electrode after the immobilization of GOD and the Nafion deposition. Figure S6. Electrochemical impedance spectroscopy (EIS) analysis of the biosensor before and after Prussian blue deposition. (a) Nyquist plot; (b) Bode plot. Figure S7. Performance of the microtube biosensor for detecting H2O2 in PBS. (a) Current-verses-time response upon the additions of different concentrations of H2O2. C1: 0.8 mM, C2: 2.2 mM, C3: 3 mM, C4: 6 mM, C5: 10 mM, C6: 12 mM, C7: 15 mM, C8: 16 mM, C9: 20 mM. Each error bar was three different biosensors. (b) Calibration curve of the biosensor for detecting H2O2. Figure S8. Optical images of the hematoxylin and eosin-stained pierced skin section after the insertion of microtube (a) and after applying a object (20 g) on the microtube. Figure S9. Camera images of the rat’s skin after application of the microtube (with or without electrode) for 5 days. Figure S10. Camera images of the rat’s skin after removing the device. Figure S11. (a) Relative error of the biosensor at different blood glucose values. (b) Distribution of the relative error. Table S1. Comparision of this work and other closed-loop diabetes management systems. (Supplementary Materials)

References

  1. P. E. H. Schwarz, G. Gallein, D. Ebermann et al., “Global Diabetes Survey--an annual report on quality of diabetes care,” Diabetes Research and Clinical Practice, vol. 100, no. 1, pp. 11–18, 2013. View at: Publisher Site | Google Scholar
  2. H. King, R. E. Aubert, and W. H. Herman, “Global burden of diabetes, 1995-2025- prevalence, numerical estimates, and projections,” Diabetes Care, vol. 21, no. 9, pp. 1414–1431, 1998. View at: Publisher Site | Google Scholar
  3. G. S. Eisenbarth, J. S. Flier, and G. Cahill, “Type I diabetes mellitus. a chronic autoimmune disease,” The New England Journal of Medicine, vol. 314, no. 21, pp. 1360–1368, 1986. View at: Google Scholar
  4. S. E. Inzucchi, R. M. Bergenstal, J. B. Buse et al., “Management of Hyperglycemia in type 2 diabetes: a patient-centered approach,” Diabetes Care, vol. 35, no. 6, pp. 1364–1379, 2012. View at: Publisher Site | Google Scholar
  5. A. Sultan, J. F. Thuan, and A. Avignon, “Primary prevention of cardiovascular events and type 2 diabetes: should we prioritize our interventions?” Diabetes & Metabolism, vol. 32, no. 6, pp. 559–567, 2006. View at: Publisher Site | Google Scholar
  6. C. Home, “Self-reported visual impairment among persons with diagnosed diabetes - United States, 1997-2010,” Morbidity and Mortality Weekly Report, vol. 60, no. 45, pp. 1549–1553, 2011. View at: Google Scholar
  7. S. Wild, G. Roglic, A. Green, R. Sicree, and H. King, “Global prevalence of diabetes,” Diabetes Care, vol. 27, no. 5, pp. 1047–1053, 2004. View at: Publisher Site | Google Scholar
  8. E. A. H. Sims and P. M. Catalano, “Response to the expert committee on the diagnosis and classification of diabetes mellitus,” Diabetes Care, vol. 21, no. 3, pp. 467-468, 1998. View at: Publisher Site | Google Scholar
  9. E. Matteucci and O. Giampietro, “Point-of-care testing in diabetes care,” Mini-Reviews in Medicinal Chemistry, vol. 11, no. 2, pp. 178–184, 2011. View at: Publisher Site | Google Scholar
  10. B. Westerhaus, A. R. Gosmanov, and G. E. Umpierrez, “Diabetes prevention: can insulin secretagogues do the job?” Primary Care Diabetes, vol. 5, no. 2, pp. 73–80, 2011. View at: Publisher Site | Google Scholar
  11. Y. W. Xu, S. J. Pilla, G. C. Alexander, and I. B. Murimi, “Use of non-insulin diabetes medicines after insulin initiation: a retrospective cohort study,” PLoS One, vol. 14, no. 2, p. 12, 2019. View at: Publisher Site | Google Scholar
  12. X. Wang, P. L. Zheng, G. Huang, L. Yang, and Z. G. Zhou, “Dipeptidyl peptidase-4(DPP-4) inhibitors: promising new agents for autoimmune diabetes,” Clinical and Experimental Medicine, vol. 18, no. 4, pp. 473–480, 2018. View at: Publisher Site | Google Scholar
  13. W. Mosleh, A. Sharma, M. S. Sidhu, B. Page, U. C. Sharma, and M. E. Farkouh, “The role of SGLT-2 inhibitors as part of optimal medical therapy in improving cardiovascular outcomes in patients with diabetes and coronary artery disease,” Cardiovascular Drugs and Therapy, vol. 31, no. 3, pp. 311–318, 2017. View at: Publisher Site | Google Scholar
  14. H. Li, T. Chang, Y. Gai et al., “Human joint enabled flexible self-sustainable sweat sensors,” Nano Energy, vol. 92, p. 106786, 2022. View at: Publisher Site | Google Scholar
  15. T. R. Chang, H. Li, N. Zhang et al., “Highly integrated watch for noninvasive continual glucose monitoring,” Microsystems & Nanoengineering, vol. 8, no. 1, 2022. View at: Publisher Site | Google Scholar
  16. L. Lin, Y. Wang, M. Cai et al., “Multimicrochannel microneedle microporation platform for enhanced intracellular drug delivery,” Advanced Functional Materials, vol. 32, no. 21, 2022. View at: Publisher Site | Google Scholar
  17. J. Y. Stone, N. Haviland, and T. S. Bailey, “Review of a commercially available hybrid closed-loop insulin-delivery system in the treatment of type 1 diabetes,” Therapeutic Delivery, vol. 9, no. 2, pp. 77–87, 2018. View at: Publisher Site | Google Scholar
  18. F. Ribet, G. Stemme, and N. Roxhed, “Real-time intradermal continuous glucose monitoring using a minimally invasive microneedle-based system,” Biomedical Microdevices, vol. 20, no. 4, 2018. View at: Publisher Site | Google Scholar
  19. Y. Q. Liu, Q. Yu, X. J. Luo, L. Yang, and Y. Cui, “Continuous monitoring of diabetes with an integrated microneedle biosensing device through 3D printing,” Microsystems & Nanoengineering, vol. 7, no. 1, 2021. View at: Publisher Site | Google Scholar
  20. C. J. Rini, E. McVey, D. Sutter et al., “Intradermal insulin infusion achieves faster insulin action than subcutaneous infusion for 3-day wear,” Drug Delivery and Translational Research, vol. 5, no. 4, pp. 332–345, 2015. View at: Publisher Site | Google Scholar
  21. S. X. Yang, F. Wu, J. Liu et al., “Phase-transition microneedle patches for efficient and accurate transdermal delivery of insulin,” Advanced Functional Materials, vol. 25, no. 29, pp. 4633–4641, 2015. View at: Publisher Site | Google Scholar
  22. O. Heifler, E. Borberg, N. Harpak et al., “Clinic-on-a-needle Array toward future minimally invasive wearable artificial pancreas applications,” ACS Nano, vol. 15, no. 7, pp. 12019–12033, 2021. View at: Publisher Site | Google Scholar
  23. H. Lee, T. K. Choi, Y. B. Lee et al., “A graphene-based electrochemical device with thermoresponsive microneedles for diabetes monitoring and therapy,” Nature Nanotechnology, vol. 11, no. 6, pp. 566–572, 2016. View at: Publisher Site | Google Scholar
  24. X. L. Li, X. Huang, J. Mo et al., “A fully integrated closed-loop system based on mesoporous microneedles-iontophoresis for diabetes treatment,” Advanced Science, vol. 8, no. 16, article 2100827, 2021. View at: Publisher Site | Google Scholar
  25. X. Luo, Q. Yu, Y. Liu et al., “Closed-loop diabetes minipatch based on a biosensor and an electroosmotic pump on hollow biodegradable microneedles,” ACS Sensors, vol. 7, no. 5, pp. 1347–1360, 2022. View at: Publisher Site | Google Scholar
  26. T. S. Tang, E. M. Digby, A. M. Wright et al., “Real-time continuous glucose monitoring versus internet-based blood glucose monitoring in adults with type 2 diabetes: a study of treatment satisfaction,” Diabetes Research and Clinical Practice, vol. 106, no. 3, pp. 481–486, 2014. View at: Publisher Site | Google Scholar
  27. A. L. Secher, A. B. Madsen, L. Ringholm et al., “Patient satisfaction and barriers to initiating real-time continuous glucose monitoring in early pregnancy in women with diabetes,” Diabetic Medicine, vol. 29, no. 2, pp. 272–277, 2012. View at: Publisher Site | Google Scholar
  28. H. Brauker James and K. A. Ullas, “Analyte Sensor,” US Patent 2015087942A1. View at: Google Scholar
  29. Z. Pu, R. Wang, K. Xu, D. Li, and H. Yu, “A Flexible Electrochemical Sensor Modified by Graphene and AuNPs for Continuous Glucose Monitoring,” in IEEE SENSORS, pp. 1–4, New York, 2015. View at: Google Scholar
  30. K. W. Johnson, J. J. Mastrototaro, D. C. Howey et al., “In vivo evaluation of an electroenzymatic glucose sensor implanted in subcutaneous tissue,” Biosensors & Bioelectronics, vol. 7, no. 10, pp. 709–714, 1992. View at: Publisher Site | Google Scholar
  31. J. Y. Zhang, T. Shang, S. Chattaraj et al., “Advances in insulin pump infusion sets symposium report,” Journal of Diabetes Science and Technology, vol. 15, no. 3, pp. 705–709, 2021. View at: Publisher Site | Google Scholar
  32. L. M. Johnson, S. A. Krovi, L. Li et al., “Characterization of a reservoir-style implant for sustained release of tenofovir alafenamide (TAF) for HIV pre-exposure prophylaxis (PrEP),” Pharmaceutics, vol. 11, no. 7, 2019. View at: Publisher Site | Google Scholar
  33. S. M. Simpson, L. Widanapathirana, J. T. Su et al., “Design of a drug-eluting subcutaneous implant of the antiretroviral tenofovir alafenamide fumarate,” Pharmaceutical Research, vol. 37, no. 4, 2020. View at: Publisher Site | Google Scholar
  34. J. E. Brown, L. Tozzi, B. Schilling et al., “Biodegradable silk catheters for the delivery of therapeutics across anatomical repair sites,” Journal of Biomedial Materials Research Part B Applied Biomaterials, vol. 107, no. 3, pp. 501–510, 2019. View at: Publisher Site | Google Scholar
  35. A. Okada, Y. Niwa, T. Katase, and O. Kai, “Controlled release of estradiol-17β and bisphenol A from a silicone tube for long-term administration in mice,” Animal Science Journal, vol. 76, no. 6, pp. 535–539, 2005. View at: Publisher Site | Google Scholar
  36. J. Madden, C. O'Mahony, M. Thompson, A. O'Riordan, and P. Galvin, “Biosensing in dermal interstitial fluid using microneedle based electrochemical devices,” Sensing and Bio-Sensing Research, vol. 29, article 100348, 2020. View at: Publisher Site | Google Scholar
  37. J. J. Norman, M. R. Brown, N. A. Raviele, M. R. Prausnitz, and E. I. Felner, “Faster pharmacokinetics and increased patient acceptance of intradermal insulin delivery using a single hollow microneedle in children and adolescents with type 1 diabetes,” Pediatric Diabetes, vol. 14, no. 6, pp. 459–465, 2013. View at: Publisher Site | Google Scholar
  38. F. R. Cichocki, G. Chen, J. Zhou, and M. Deng, “Synthetic bioabsorbable microtubes for the treatment of surgical sites,” Materials Science and Engineering: C, vol. 28, no. 3, pp. 331–335, 2008. View at: Publisher Site | Google Scholar
  39. K. Ngamchuea, S. Eloul, K. Tschulik, and R. G. Compton, “Planar diffusion to macro disc electrodes-what electrode size is required for the Cottrell and Randles-Sevcik equations to apply quantitatively?” Journal of Solid State Electrochemistry, vol. 18, no. 12, pp. 3251–3257, 2014. View at: Publisher Site | Google Scholar
  40. G. Leftheriotis, S. Papaefthimiou, and P. Yianoulis, “Dependence of the estimated diffusion coefficient of Li _x_ WO3 films on the scan rate of cyclic voltammetry experiments,” Solid State Ionics, vol. 178, no. 3-4, pp. 259–263, 2007. View at: Publisher Site | Google Scholar
  41. A. Arjunan, A. Sukeri, D. P. M. Saraiva, P. B. Miranda, and M. Bertotti, “Electrochemical studies of hydrogen peroxide oxidation on a nanoporous gold surface: fundamental and analytical applications,” Journal of the Electrochemical Society, vol. 167, no. 11, 2020. View at: Google Scholar
  42. K. K. Wu and Y. Huan, “Streptozotocin-induced diabetic models in mice and rats,” Current Protocols in Pharmacology, vol. 5, pp. 45–47, 2008. View at: Google Scholar
  43. A. Pfutzner, D. C. Klonoff, S. Pardo, and J. L. Parkes, “Technical aspects of the Parkes error grid,” Journal of Diabetes Science and Technology, vol. 7, no. 5, pp. 1275–1281, 2013. View at: Publisher Site | Google Scholar
  44. J. C. Yu, J. Wang, Y. Zhang et al., “Glucose-responsive insulin patch for the regulation of blood glucose in mice and minipigs,” Nature Biomedical Engineering, vol. 4, no. 5, pp. 499–506, 2020. View at: Publisher Site | Google Scholar
  45. E. Renard, B. Guerci, A. M. Leguerrier, and R. Boizel, “Lower rate of initial failures and reduced occurrence of adverse events with a new catheter model for continuous subcutaneous insulin infusion: prospective, two-period, observational, multicenter Study,” Diabetes Technology & Therapeutics, vol. 12, no. 10, pp. 769–773, 2010. View at: Publisher Site | Google Scholar